Pub. # OM20-MSP3-3D
4-22
Figure 42
2. Adjust electrode holder: with sampling mechanism raised completely, adjust electrode
holder to provide ½” (1.2 cm) clearance between it and sampling trough.
Figures 43a & b
3. Wash adjustment: Wash brackets should be parallel to sampling trough, with jets directly
beside electrode holder, jets should be 1” (2.5 cm) below electrodes; when electrodes are
properly aligned.
Figures 44a & b
Turn crank to raise
or lower electrode
holder
electrode
holder 1”
(2.5cm)
above wash
jets
Adjust scraper bracket
until cutting shoe
clears scraper when
sampler assembly is
raised completely.
½” (1.2 cm) clearance
between electrode holder
and trough liner