Vantage 5000
1.4v-09.2019
28 | Page
Figure 23 Calibration Menu
7.5 Calibration
To access the calibration menu, touch the icon
as indicated in
Figure 8
, item 2.
1. QUANTITY (QTY) – sets the calibration quantity to be
dispensed.
2. JOB FILE NAME – displays the current Activated Job File
for Calibration. It is a read-only display and can only be
changed in
(
g
7.3 Job Files)
.
The calibration is applicable for the displayed Job File only. The
speed control and/or start/stop control must be in HMI mode
(
g
7.8 Remote Control).
3. UOM (UNIT OF MEASURE) – sets the unit of measure
from a drop-down list for the quantity dispensed.
a. ml (Millilitre)
b. Grams
c. Litres
d. Pounds
e. USG (US Gallons)
4. PRODUCT SG (SPECIFIC GRAVITY) – displays the
product specific gravity from the Activated Job File. It is
a read-only value and can only be changed in
(
g
7.4.1
Delivery Setup).
5. FLOW RATE & CALIBRATION TIME INDICATOR
u
FLOW RATE – displays the flow rate from the
Activated Job File. This is a read-only value and can
only be changed in
(
g
7.4.1 Delivery Setup).
u
CALIBRATION TIME – displays the estimated time in
seconds for the set quantity to be dispensed.
2
3
4
5
6
6. JOB NOT CALIBRATED
– will be displayed if NO
calibration has been performed for the Activated Job File.
When the calibration has been performed,
a message will
be displayed as follows: “
Last calibrated DD/MM/YYYY at
HH:MM:SS [x.xx]
”.
7. GO BACK/CANCEL – cancels any changes and return to
the Job File Menu. If the user has made changes, there
will be a prompt asking the user if the changes should be
saved.
8. HOME – returns the user to the Home Screen. If the user
has made changes, there will be a prompt asking if the
changes should be saved.
1
7
8
Note