Operations & Maintenance Manual: Important Information
The docking station, charging cradle, power adapter, and power cords are not intended for
patient contact. Ensure 2 m (6 ft) is maintained between the patient and these components.
WARNING
Ensure proper distance from patient. When transmitting data to or from your computer, make sure
the device, accessories, and computer are outside the patient vicinity (more than 2 m (6 ft) from
the patient).
WARNING
Do not use the system on:
• Fetal patients.
• Pregnant patients.
• Patients with open skin or wounds in the suprapubic region.
• Patients with ascites.
WARNING
To reduce the risk of electric shock or burns, do not use the system in conjunction with
high‑frequency surgical equipment.
WARNING
To reduce the risk of electrical shock, do not attempt to open the system components. This may
cause serious injury to the operator or damage to the device and will void the warranty. Contact
Verathon
®
Customer Care or your local representative for all servicing needs.
WARNING
Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and cause improper operation.
WARNING
Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30cm (12inches) to any part of the
BladderScan BVI 6400 system, including cables specified by Verathon. Otherwise, degradation
of performance of this equipment could result.