24
PROCEdURE 1. CLEAN & dISINFECT THE dEVICE
Use this procedure in order to clean and disinfect the device. Review the information in the
and
sections before completing this procedure.
CLEAN THE dEVICE
1. Put on new gloves.
2. After every exam, using a dry paper towel or soft cloth, wipe any ultrasound gel completely off the scanhead.
3. Use a soft, moistened cloth to remove particulate matter or body fluids that remain on the device.
4. Allow the device to air dry or towel dry with a clean dry cloth before disinfecting.
5. Continue to the following section,
. Low‑level disinfection of the scanhead is required
between uses.
dISINFECT THE dEVICE
Low‑level disinfection of the scanhead is required between uses. Only use disinfectants prior to their
expiration date.
6. Remove the gloves used in the cleaning portion of the procedure, and then put on new gloves.
7. If using a liquid disinfectant, mix the disinfection solution according to the manufacturer’s label
instructions for the appropriate disinfection level concentration.
8. Apply the solution to a soft cloth or wipe for application. Do not spray or apply liquid disinfectants
directly to the surface of the device or soak the device in liquids.
9. Wipe the surfaces of the device, allowing the surface to stay wet for the required contact duration.
Follow the manufacturer’s instructions for the appropriate disinfection level contact duration.
10. If rinsing or removal of the disinfectant solution from the device is required by the disinfectant
manufacturer’s instructions, wipe the device with a clean soft cloth dampened in sterile water. Verathon
®
recommends wiping the device three separate times to remove all residual disinfectant.
11. Allow the device to air dry or towel dry with a clean, dry cloth.