ELGA
MEDICA-D 7/15 US
Operator Manual
MEDICA-D 7/15 US MANU40550 Version 2
– 05/17
Page 7
3.3 Technical Specifications
The specifications for the
MEDICA-D 7/15 US
are as follows:
Feedwater
MEDICA-D 7 US
MEDICA-D 15 US
Feedwater
Source quality
Potable mains water supply
Potable mains water supply
Fouling index-maximum
10
10
Total dissolved solids-maximum
1400
S/cm
1400
S/cm
Free chlorine - maximum
0.5ppm
0.5ppm
Heavy metals - maximum
0.05ppm
0.05ppm
Silica- maximum
30ppm
30ppm
Temperature
1 - 35°C
1 - 35°C
Flowrate
(maximum requirement)
78 l/hr
85 l/hr
Feedwater Pressure
Maximum - without internal boost pump
6.0 bar (90 psi)
6.0 bar (90 psi)
Minimum - without internal boost pump
4.0 bar (60 psi)
4.0 bar (60 psi)
Maximum - with internal boost pump
2.0 bar (30 psi)
2.0 bar (30 psi)
Minimum - with internal boost pump
Flooded suction
Flooded suction
Drain requirements (gravity fall with air gap). Maximum during service.
70 l/hr
70 l/hr
Connections
Inlet - quick connect
8mm (5/16") OD
8mm (5/16") OD
Outlet - quick connect
8mm (5/16") OD
8mm (5/16") OD
Drain RO - quick connect
8mm (5/16") OD
8mm (5/16") OD
Analyzer feed
8mm (5/16") OD
8mm (5/16") OD
Delivery pump feed - quick connect
8mm (5/16") OD
8mm (5/16") OD
Positioning
Wall, bench or under bench
mounted.
Wall, bench or under bench
mounted.
Environment
Clean dry indoor.
Temp 5 - 40°C.
Humidity max 80%
non-condensing.
Clean dry indoor.
Temp 5 - 40°C.
Humidity max 80%
non-condensing.