Water Purity Alarm
This alarm will signal if the water purity deviates from the preset
parameters and will cause the water purity value to flash and an
alarm to sound, until water purity improves to within acceptable
purity limits. Press the
button to mute the alarm. If water
purity stays outside acceptable purity limits replace the Ion-
exchange pack following the instructions in
Section 7.2.
Reservoir Disconnect Fault Alarm
The reservoir level disconnect fault alarm condition will signal
with an audible alarm and flashing icon. PRESS the button to
mute the alarm. This alarm condition will cause the process to
turn off. Connect the reservoir level lead and power the unit off
and on to clear the alarm.
Vacuum Fail Alarm
The vacuum fail alarm will signal with an audible alarm and
flashing icon of a circular vacuum gauge. This may indicate
blockage of the ejector or failure of the degassing membrane.
Flash
Water Purity Alarm
Reservoir Level
Disconnect Alarm
Vacuum Fail
Alarm