Operator Manual
MEDICA-D 7/15 US MANU40550 Version 2
– 05/17
Page 17
Step 4 - Water Purity Unit Setting
This screen allows preferred water purity unit of measure to be
set, to either,
M
.cm
or
µS/cm
. Once selected, all future water
purity measurements will be displayed with your unit of choice.
SELECT the option required using the
button and ACCEPT
with the
button.
Step 5 - Pump Speed Setting
Adjustment of this setting should only be made under specific
instructions from ELGA LabWater Technical Support.
1. Press
to accept.
Note:
To reset any of the set-up parameters,
restart from the set up menu and follow
instructions from Step 1.
Step 6 - Purity Alarm Setting
This screen is used for setting the value at which the water
purity alarm will activate.
If the
M
.cm
water purity unit setting was chosen, then the
following water purity alarm choices will be displayed.
15 M
.cm
10 M
.cm
1 M
.cm
If the
µS/cm
water purity unit setting was chosen, then the
following water purity alarm choices will be displayed.
0.067 µS/cm
0.100 µS/cm
1.0 µS/cm
SELECT the alarm setting required using the
button and
ACCEPT with the
button.
Megaohm Microsiemen
Purity Alarm Setting
Megaohm
Megaohm
Microsiemen
Water Purity Settings
100%
Pump Speed Setting