ELGA
MEDICA-D 7/15 US
Operator Manual
MEDICA-D 7/15 US MANU40550 Version 2
– 05/17
Page 33
Step 1 - Start Sanitization Cycle
1. ENSURE that the docking vessel/reservoir level
indication on the graphics display is showing
>10%. If display shows >40%, >70% or 100%,
dispense water until display changes to >10%.
2. PRESS the PROCESS button to stop the
process.
3. TURN OFF the electrical supply.
4. WAIT one minute for residual pressure to
dissipate.
5. ISOLATE the pump feed line from the reservoir
by closing valve V3 (see page 15) then
disconnect the tube from the unit side of the valve
(V3). Direct the line to drain.
Step 2
– Pour sterilant into Sanitization Port
1. UNSCREW cap on sanitization port.
2. MEASURE out 20ml of Minncare Cold Sterilant
and SLOWLY POUR into sanitization port.
Note: It is recommended that you add the
chemical in 10ml steps. If the chemical level rises
too high in the sanitization port, refit the cap and
then remove it. This will allow room for the
remaining 10ml.
3. REFIT cap on sanitization port, hand tight.
Step 3 - Start the Sanitization Process
1. RESTORE the electrical supply.
2. PRESS the PROCESS button.
3. ALLOW the sanitization and rinsing to continue
for 2 hours.
4. PRESS the PROCESS button to stop the
process.
5. TURN OFF electrical supply.
Step 4 - Return to Normal Operation
1. Reconnect the pump feed to valve V3 and open
the isolating valve (V3).
2. TURN ON the electrical supply.
3. PRESS PROCESS button to return to normal
operation.
4. CHECK for leaks.
8.1.2
Sanitization Procedure for the unit and docking
vessel/reservoir
The unit and docking vessel/reservoir may be sanitized to destroy
bacteria within the pipework and filters of the unit. The complete
sanitization of the system is recommended to be performed if high
levels of bacterial contamination are found in the product water
feeding the analyzer. Should bacterial contamination of the product
water be suspected, samples should be taken by trained staff and
analyzed for bacterial counts. If the counts are abnormally high, ELGA
LabWater technical support or your local representative should be
contacted for advice and assistance.