PART 3 - Technical manual
34
DISMOUNTING, STORAGE AND REMOUNTING
Dismounting
VENDLET Basic may be dismounted and transferred to
another bed.
1. Remove the turning sheet and slide sheet.
2. Raise the VENDLET Basic bars to highest position.
3. Disconnect the power cable from the outlet.
4. Disconnect the cables from the control unit.
5. Dismount the control unit by cutting cable ties.
6. Now VENDLET Basic can be dismounted. Loosen the bolts
in the mounting plates and turn them aside to lift VEND-
LET Basic off the bed frame. Optionally, use the mounting
trolley to transport VENDLET Basic to a new destination.
Storage
Store VENDLET Basic in the original packaging until mounting.
Indoor
0
+50
0° C (32° F) Min
+50° C (122° F) Max
20
90%
20–90% at 30°C
Keep dry
Remounting
Follow the procedure described in the section “Mounting” if
remounting is required.
Before remounting, you should check all parts of VENDLET
Basic for damage and that the date of the annual service check
has not been exceeded, and that VENDLET Basic has been
properly cleaned / disinfected.