Velp Scientifica
– UDK139 Operating Manual
10004762/B5
41
To return to the "System" menu press the arrow top left.
Select Upload software. The display shows:
The software can be upgraded by following a few simple, guided steps.
Insert the USB pen drive containing the updated software file into the USB port of
the instrument.
A green flag on the display indicates that the software was uploaded correctly, a red
X indicates that the procedure failed. In this case check that the USB pen drive is in
good working order, that it is inserted properly in the USB socket and that the name
of the file is
UDK139_update.zip
.
Follow the instructions on the display to complete the upgrade procedure.
If there is a power failure during upgrade, restore the electricity and turn the
instrument off for 5 seconds. Turn the instrument on again and check the software
version shown at the bottom of the display to make sure that the upgrade was
uploaded successfully.
228