Velp Scientifica
– UDK139 Operating Manual
10004762/B5
12
3.
Assembly and installation
After unpacking the instrument position it on a laboratory bench or on a non-
flammable surface at a distance of at least 30 cm from the walls so as not obstruct
the air vents on the rear.
NOTE:
The instrument weighs 31kg. If you need to move the instrument suitable
lifting devices should be used.
Do not remove the protective film from the touch screen.
Position the instrument near the socket power supply because it is
considered a means of disconnecting the device
The instrument must be placed near a water tap (necessary for cooling the distillate)
with a maximum pressure of 5 bar, and sink in order to discharge the waste water.
The instrument requires a tap water flow rate (depending on the temperature of the
tap water) of approximately:
- 0.5 l/minute at a temperature of 15°C
- 1 l/minute at a temperature of 30°C
Water is required during the distilling phase only and is interrupted automatically
during the other phases.
Position the
tanks for NaOH and H
2
O solutions and for collecting distillation residues
below or behind the unit. The tanks are not supplied with the instrument because
their volume is chosen according to specific requirements. Usually polythene carboys
of 5, 10 or 20l are used.
The hydraulic tubes are supplied with the
instrument and the hydraulic connections
are clearing identified on the rear of the
instrument.