Velp Scientifica
– UDK139 Operating Manual
10004762/B5
38
4.6.2. Calibration
Select Calibration from the
“System” menu.
The following sub-menus are listed:
To return to the "System" menu press the arrow top left.
The "Calibration" menu is used to calibrate the flow rate of the reagent pumps.
NOTE:
Make sure that the hydraulic circuits are filled before calibrating (see "Check-
up").
Select H
2
O pump calibration. The display shows:
Press
to begin the process. The instrument measures 50ml of diluting water.
Transfer the reagent (diluting water) to a graduated cylinder to measure the real
volume. Use the numeric keypad to enter the real volume, press
confirm.
Repeat the procedure for the NaOH pump.
225