Cardio-M Plus 

OPERATION 

MANUAL 

 

 

 

 
 
 

 

 

Summary of Contents for Cardio-M Plus

Page 1: ...1 83 Version 1 0 Cardio Cardio Cardio Cardio M Plus M Plus M Plus M Plus Operation Manual Operation Manual Operation Manual Operation Manual Cardio M Plus OPERATION MANUAL ...

Page 2: ...ies runs for 6 months Warranty repair or replacement will be made by the Medical Econet GmbH Service Center at no charge for the warranty period if properly used under normal conditions in accordance with the instructions for use In the event of a malfunction or failure during the warranty period the customer should inform Medical Econet GmbH of the model name serial number date of purchase and a ...

Page 3: ...the telephone number or e mail address below You can talk to our sales people Medical Econet GmbH always welcomes your inquiries Please contact us Customer service 49 0 2365 92437 0 When the product is not operating correctly please contact our customer service center with the name of the model serial number and a description of the problem you are experiencing Technical assistance Please contact ...

Page 4: ...over cleaning adjusting and instructions for use 2 A fault of customer When the product is damaged or broken due to customer s improper use or misuse When the power cord is replaced improperly by the user When the user drops the product while moving When the user uses unauthorized accessories When the product is repaired by an unauthorized person 3 Others When the product is damaged by any externa...

Page 5: ...ing and caution instructions The manufacturer or service agents are not responsible for the damage resulting from inappropriate use or carelessness WARNING There may be serious injuries fatal accidents or financial damages if you violate these instructions CAUTIONS There may be slight injuries or reduced damages if you violate these instructions NOTE There may not be any dangerous events but it is...

Page 6: ...mperature ranges from 10 to 40 C 50 to 104 and humidity ranges from 30 to 85 Avoid placing in the vicinity of electric heaters Avoid placing in an area where there is an excessive rise in humidity or a ventilation problem Avoid placing where there is an excessive shock or vibration may occur Avoid placing in an area where chemicals are stored or where there is danger of gas leakage Avoid inserting...

Page 7: ...ith other appliances the result may also be inaccurate NOTE The classification of Cardio M Plus is as below Its classification against electric shocks is Class Ⅰ Type BF It is not proper to use this product near a flammable anesthetic or solvent IEC EN 60601 1 Safety of Electric Medical Equipment standard noise events A class IEC EN60601 1 2 Electromagnetic Compatibility Requirements standard nois...

Page 8: ...Plus M Plus Operation Manual Operation Manual Operation Manual Operation Manual Safety Symbols Symbols Contents Isolated patient connection IEC 601 1 Type BF Power on Power off ATTENTION Refer to the user s manual Ground connection External Signal IN OUT Port ECG ...

Page 9: ...nd Accessories 14 Optional Components 15 Body Configuration 16 Front Panel 19 Control Panel 21 Power 22 4 System Installation 24 Precautions For Installation 24 Power Connection 24 Patient Cable Connection 24 Paper Installation 24 System Start 25 5 System Setup 26 Reset 26 Chapter 2 Preparation For Electrocardiography 27 1 Location Of Electrode 28 2 Electrode Connection 29 Patient Cable Connection...

Page 10: ...inting Setup 37 Grid On Off 37 Beat Report Setup 38 Diagnosis Printing Setup 39 Rhythm Lead 39 Patient Information Input 40 Pediatrics Diagnosis 42 3 LCD Monitor Display 43 4 Rhythm Mode Printing 45 Printing Speed 45 How To Print 45 5 Record Mode Printing 46 10 Second ECG Recording 46 60 Second ECG Recording 47 5 Minute ECG Recording 48 6 Printing Form 49 7 Copy Mode Printing 55 8 Device Setup 57 ...

Page 11: ...io Cardio M Plus M Plus M Plus M Plus Operation Manual Operation Manual Operation Manual Operation Manual Chapter 4 System Management 75 1 Maintenance And Cleaning 76 2 Regular Check Up 76 3 Trouble Shooting 76 Chapter 5 Specification 78 ...

Page 12: ... Chapter 1 General Information 1 Product Overview 2 Product Features 3 Product Configuration Basic Components And Accessories Optional Components Body Configuration Front Panel Control Panel Power 4 System Installation Precautions For Installation Power Connection Patient Cable Connection Paper Installation System Start 5 System Setup Reset ...

Page 13: ...ttery optional can be added to the device so that the device can be operated conveniently during bedside visits or emergency conditions 2 Product Features Print 12 channel ch cardiac signals with reports in 3ch 1 rhythm 3ch 3 rhythms 6ch 1 rhythm and 12ch rhythm patterns on A4 size printing paper Record a rhythm from 1 channel for 60 seconds or 5 minutes then print it on A4 size printing paper Rec...

Page 14: ...ion The Cardio M Plus system consists of the items below Unpack the package and check the items below are included Also be sure to check for any damage to the body and accessories Basic Components And Accessories Cardio M Plus Body 1 ea Patient Cable 1 ea Disposable Electrodes 1 set ECG Clips 1 set ECG Paper 1 ea Rechargeable Battery User Manual 1 ea Power Cord ...

Page 15: ...Cardio Cardio Cardio Cardio M Plus M Plus M Plus M Plus Operation Manual Operation Manual Operation Manual Operation Manual Body Configuration Top View Handle Printer Cover Printer Cover Opening Button LCD Control Panel ...

Page 16: ...lus M Plus M Plus M Plus Operation Manual Operation Manual Operation Manual Operation Manual Front View Printer Cover Opening Button Rear View Protective Ground Terminal Power Switch USB Port AC Power Connection Port RS 232C Serial Port RJ45 LAN Port ...

Page 17: ...ion Manual Operation Manual Operation Manual Operation Manual Left Side View Handle Right Side View Patient Cable Connection Port NOTE To avoid an expected electric shock do not open the equipment cover or disassemble the equipment Refer servicing to Medical Econet GmbH America Inc ...

Page 18: ...dio Cardio Cardio M Plus M Plus M Plus M Plus Operation Manual Operation Manual Operation Manual Operation Manual Front Panel Rotary key Power status Battery status Graphic display window Rhythm Record Copy ECG Spiro View Net ESC ...

Page 19: ... 1 0 Cardio Cardio Cardio Cardio M Plus M Plus M Plus M Plus Operation Manual Operation Manual Operation Manual Operation Manual During device boot up you can see the system version and the Medical Econet GmbH name ...

Page 20: ...ys Patient ID Displays The Date Displays Battery Status or AC Power Connection Status Displays ECG Sensitivity with One of 5 10 20 and Auto mm mV Displays Printing Speed With One Of 12 5 25 And 50 mm sec Displays Number Of Data In Memory Displays A Printing Report Channel Form With One Of 3ch 1rhy 3ch 3rhy 6ch 1rhy 12ch rhy 60s 1rhy and 5m 1rhy Displays a Rhythm Channel Setup with One Of I II III ...

Page 21: ...paper Print out the data in a selected form after recording 10 sec data or 60 sec data Process the recorded data with or without revision and print them out Change a view of the graphic display window to initial main window Change the view of the graphic display window NET Execute communication using networking ESC Cancel commands or change into the former mode for main menu printing This key is c...

Page 22: ...ed by touching menu items on the screen Power AC Power When AC Power is connected to the device power LED light will be green if the battery is installed charging will commence AC Power Battery Power Battery Power The device will get its power from the battery and the unit will display the battery power icon shown above when the system is powered on the battery is connected to the device and the A...

Page 23: ... PACK Ni MH battery 12V 2600mAh When to replace Battery will automatically be charged when the device is connected to AC power and cannot be charged when separated from the device Battery is designed to have a charging cycle of 500 times or more If the device could only last 20 minutes or less on a battery power the battery needs to be replaced Additionally when a battery pack is damaged or leakin...

Page 24: ...e will be recorded in the internal memory even when it is off Prevent any shock or excessive force that may cause damage to the device Place the device away from any dust or flammable materials Power Connection The equipment needs electrical power to operate Plug in one end of the power cable to wall socket and the other to Cardio M Plus Patient Cable Connection Connect the patient cable to the pa...

Page 25: ...l System Start After all installation processes are finished turn on the power switch to see the Cardio M Plus logo system firmware version the company name and icons for ECG or spirometer mode displayed on the LCD screen To select ECG mode locate the selection box on the ECG and press the rotary key or touch the selection box on the screen ...

Page 26: ...d and then reset starts SYSTEM MESSAGE Factory Default Setting Loading Defaults Are As Follows Sensitivity SENS 10mm mV Printing speed SPD 25mm s Grid GRID Off Channel form FORM 6ch 1rhy Rhythm channel RHY II Base Line filter BASE On EMG filter MUSC Off Power noise filter AC 60Hz when ENG ENG_US setting 50Hz for other languages Low pass filter LPF 40hz Diagnosis Diagnosis On Stored Data List Delet...

Page 27: ... Plus M Plus Operation Manual Operation Manual Operation Manual Operation Manual Chapter 2 Preparation For Electrocardiography 1 Location Of Electrode 2 Electrode Connection Patient Cable Connection How To Attach Electrodes Countermeasures For Bad Lead Connection ...

Page 28: ...V5 V6 as shown below Limb electrodes are located as follows RL N Right Leg LL F Left Leg RA R Right Arm LA L Left Arm V1 C1 Fourth Intercostal Space At The Right Border Of The Sternum V2 C2 Fourth Intercostal Space At The Left Border Of Sternum V3 C3 Midway Between Location V2 and V4 V4 C4 At The Mid Clavicular Line In The Fifth Intercostals Space V5 C5 At The Another Axillary Line On The Same Hor...

Page 29: ...t Cable Connection Connect the patient cable to the patient cable port of the Cardio M Plus How To Attach Electrodes Attach electrodes after placing the patient on the bed helping him her to be relaxed and cleaning electrode contact points on the patient s skin by using a disinfective alcohol or water Finally attach leads to the parts as explained in aforementioned location of electrode part Cauti...

Page 30: ...t to ON in System Setup and will NOT be displayed when set to OFF Lead Fault message will appear and a buzzer alarm will go off when lead fault occurs during printing The following two cases could cause a lead fault error First case leads are detached from the patient s body In this case re attach leads normally Second case signal conductivity between leads and the patient s body is low In this ca...

Page 31: ... Setup General Sensitivity Setup Printing Speed Setup Channel Form Setup Grid On Off Rhythm Printing Setup Beat Report Setup Diagnosis Printing Setup Rhythm Lead Patient Information Input Pediatrics Diagnosis 3 LCD Monitor Display 4 Rhythm Mode Printing Printing Speed How To Print 5 Record Mode Printing 10 Second ECG Recording 60 Second ECG Recording 5 Minute ECG Recording 6 Printing Form ...

Page 32: ...dio Cardio Cardio M Plus M Plus M Plus M Plus Operation Manual Operation Manual Operation Manual Operation Manual 7 Copy Mode Printing 8 Device Setup Basic Setup Network Setup Hospital Setup 9 Network Functions 10 Data Management ...

Page 33: ...he patient To copy the same report as the previously printed report press the COPY key To do monitoring signal by printer press the RHYTHM key To stop the machine during the printing or saving processes press the STOP key The following buttons will be most commonly used when operating Cardio M Plus Rhythm Printing Pause monitoring on the LCD display and print real time waveform using the printer C...

Page 34: ...nu Mode Operate by rotating the rotary key clockwise or counter clockwise Menu Navigation Operate by rotating the rotary key Menu movement will occur in the same direction as the direction of the key s rotation Menu Selection Initiate by pressing the rotary key Menu box for setting of the selected menu will be displayed Modifying setting values Move to desired setting then modify select values in ...

Page 35: ...elect the desired sensitivity level and press down to enter or select it using the touch screen to set the size of waveform displayed on screen Selected values are shown on the bottom of EKG printouts in the following way when 2 5 5 10 and 20 are selected they will be displayed as all channels 2 5mm mV all channels 5mm mV all channels 10mm mV and all channels 20mm mV When Auto is selected sensitiv...

Page 36: ...ize to 25mm s Channel Form Setup Select the channel form setting of the Printing Form Currently 6 types of channel forms 3ch 1rhy 3ch 3rhy 6ch 1rhy 12ch rhy 60s 1rhy and 5m 1rhy are offered in Cardio M Plus To select the printout channel form settings place the selection block to the print setup menu and press down using the rotary key or touch screen input to select the desired channel forms in t...

Page 37: ...then V1 V2 V3 V4 V5 V6 for the last 5 seconds Additionally it will display the rhythm of the selected 1 lead on the bottom for 10 seconds 12ch rhy mode will record 10 second EKG measurements of all 12 leads in the order of channels I II III aVR aVL aVF V1 V2 V3 V4 V5 V6 60 sec 1rhy mode will record the EKG of the selected 1 lead for 60 seconds with 6 lines of 10 second long measurements 5 min 1rhy...

Page 38: ...unit is OFF Beat Report Setup During the diagnostic print set the output of the representative beat information The beat report is added to regular printout 3ch 1rhy 3ch 3rhy 6ch 1rhy and 12ch rhy output and has Representative Beat Report printed on the upper part of the printout form This Beat report has representative waveforms of each lead and shows any important parameters calculated from reco...

Page 39: ...he selection block to the Rhythm Lead setting located at the bottom of the screen and select using either the rotary key or touch screen to enter the Rhythm Lead setting In the Rhythm Lead setting select the desired channel then select the desired rhythm lead using the rotary key or touch screen The lead selected from channel 1 will become the base lead in 3ch 1rhy and 6ch 1rhy modes showing on th...

Page 40: ...al Patient Information Input In this menu operators can input the patient s ID name age sex height weight race and smoking habit In order to enter the patient s information move the selection block to the Patient Information input setting and select input fields using either the rotary key or touch screen Patient Information Selection Mode ...

Page 41: ...er selecting the value to enter using either the rotary key or touch screen Use the rotary key or touch desired values on the screen to enter information Input mode includes Keyboard for both characters and numbers and Keypad mode for numbers only On values where only the number should be entered users will see the keypad appear instead of the regular keyboard input Information Input Mode ...

Page 42: ...diatric Diagnosis Pediatric diagnosis is supported for patients ages 3 to 15 years old When entering patient information if the input age entered by the operator in the Age section falls between 3 to 15 years old the device will automatically apply a pediatric diagnosis If the patient is 16 years old or older the device will perform a regular adult diagnosis instead ...

Page 43: ...ram Filter EMG Off fixed Power Supply Filter AC Option Low Pass Filter LPF 40hz fixed The printing speed can be selected to 12 5 mm s 25 mm s and 50 mm s and the sensitivity can be selected to 2 5 mm mV 5 mm mV 10 mm mV 20 mm mV and Auto Display output can either be done in portrait or landscape views If it s in portrait mode press the VIEW key to switch viewing mode There are a total of 3 differe...

Page 44: ... M Plus M Plus M Plus Operation Manual Operation Manual Operation Manual Operation Manual Note Display output mode can be set in Basic Setup under the Device Setup menu Please refer to the Device Setup Displayed heart rate is the average of 4 heartbeats ...

Page 45: ...ogram Filter EMG Off Power Supply Filter AC Option Low Pass Filter LPF 40hz Printing Speed Printing speed can be set to 12 5 mm s 25mm s 50mm s and signal level can be set to 2 5mm mV 5mm mV 10mm mV 20mm mV and Auto In rhythm mode printing it prints only to 25mm s if you select the A4 printing How To Print The Filter setting value is fixed Set the required printing speed and sensitivity and press ...

Page 46: ...ected for the channel form press the RECORD key to display the saving time as below and save the 10 second data to the memory SYSTEM MESSAGE Record 10 seconds When the 10 second data saving is completed the following message will be displayed on the LCD screen and the setup filter will be applied to the saved data After that the parameters that were required for the diagnosis will be extracted fro...

Page 47: ...SAGE Record 60 seconds Once the EKG recording is done the following message will appear on the LCD display while recorded data is filtered and processed to acquire the heart rate SYSTEM MESSAGE Data Processing After data processing is finished the following message will appear on the LCD screen followed by printing of the recorded signal SYSTEM MESSAGE Printing Press the PAUSE key to pause or exit...

Page 48: ...ecord EKG data while printing record duration for 5 minutes on a predetermined rhythm channel Once the EKG recording is done the following message will appear on the LCD screen followed by printing of the recorded signal SYSTEM MESSAGE Printing Press the PAUSE key to pause or exit during data saving or printing After pressing the key the following message will be displayed on the LCD screen for 1 ...

Page 49: ...rdio Cardio Cardio M Plus M Plus M Plus M Plus Operation Manual Operation Manual Operation Manual Operation Manual 6 Printing Form In this section descriptions and samples of various printing forms are provided Rhythm Report Form ...

Page 50: ...50 83 Version 1 0 Cardio Cardio Cardio Cardio M Plus M Plus M Plus M Plus Operation Manual Operation Manual Operation Manual Operation Manual Record Report Form 3ch 1rhy Record Report Form 3ch 3rhy ...

Page 51: ...51 83 Version 1 0 Cardio Cardio Cardio Cardio M Plus M Plus M Plus M Plus Operation Manual Operation Manual Operation Manual Operation Manual Record Report Form 6ch 1rhy Record Report Form 12ch rhy ...

Page 52: ...0 Cardio Cardio Cardio Cardio M Plus M Plus M Plus M Plus Operation Manual Operation Manual Operation Manual Operation Manual Record Report Form 60 sec 1rhy Record Report Form 5 min 1rhy Record Report Form Beat Report ...

Page 53: ...53 83 Version 1 0 Cardio Cardio Cardio Cardio M Plus M Plus M Plus M Plus Operation Manual Operation Manual Operation Manual Operation Manual ...

Page 54: ...ration Manual Operation Manual Beat Report Parameters PR PR Interval PA P Amplitude PD P Duration QA Q Amplitude QD Q Duration RA R Amplitude RD R Duration SA S Amplitude SD S Duration QLRD Duration Between Q And Last R QRSD QRS Duration QRSA QRS Amplitude TA T Amplitude STA ST Amplitude QTc Collect Q T Interval ...

Page 55: ...message will be displayed on the LCD screen and the printing process will start SYSTEM MESSAGE Printing If Copy Mode Printing is called without performing an EKG recording there would be no EKG data to copy In this case the following message box will be displayed for 1 second followed by the setup status of the system being displayed SYSTEM MESSAGE No ECG recorded Note If the original record time ...

Page 56: ... encounters a problem before it finishes printing the following message will be displayed for 1 second before returning to previous menu SYSTEM MESSAGE Abnormal End Press the PAUSE key in order to abort the printing Once printing is stopped the following message will be shown for one second on the LCD screen before returning to the previous menu SYSTEM MESSAGE Process Canceling ...

Page 57: ...Cancel button to abandon the setting To exit from Device Setup press the NET ESC key on the control panel or select Exit section Basic Setup Basic Setup consists of Date Time Touch screen Touch Setup Unit Language Save Screen Filter QRS P M L F Line Demo and QikPrint menus Use the rotary key or touch screen to select a desired menu item which would then open its submenu to allow the user to enter ...

Page 58: ...al Operation Manual Operation Manual Operation Manual Setup Date And Time You can set the month date hour minute and second information using the rotary key in Date Time menu Touch screen functionality update planned Touch Setup You can calibrate the touch screen by setting the coordinates ...

Page 59: ... Unit You can set units of height and weight information in Cardio M Plus Currently users can choose between metric and U S standard units cm Kg and in lb Language You can set up the language for displaying and printing in Cardio M Plus You can choose languages among ENG English ENG_US Capital letters printing CHI Chinese GER German FRE French RUS Russian SPA Spanish ITA Italian TUR Turkish POL Po...

Page 60: ...ual Operation Manual Operation Manual Saving In this menu users can decide whether a message box to ask the user about saving data should appear or not If ON is selected there will be a message box asking users about saving data If OFF is selected the message box will not appear and no data will be saved ...

Page 61: ...dio Cardio M Plus M Plus M Plus M Plus Operation Manual Operation Manual Operation Manual Operation Manual Screen Mode Setup For screen mode users can choose either portrait or landscape layouts Portrait Mode Landscape Mode Filter Setup ...

Page 62: ...elect it using the rotary key or by touching the menu on the touch screen Upon selection the Filter Setup menu shown below will appear Use either the rotary key or touch screen to navigate and select menu items to edit After editing OK must be selected in order to apply the changed value If the recorded EKG signal contains a lot of noise Low Pass Filter LPF may help to clear it For Low Pass Filter...

Page 63: ...ould be no AC noise being generated In this case please set the AC filter setting to OFF so that output signal would not be distorted The status of the AC filter can be checked on the bottom part of the printout indicating AC 60Hz AC 50Hz or AC OFF Base line drift is caused by the patient s breathing It can be seen as ECG signals being overlapped on a large parabola To remove Base line drift from ...

Page 64: ...iginal output Therefore you may reduce the possibility of signal distortion if you set the filters OFF when the signals are clear For your reference we recommend that Baseline and AC filters be set to ON most of the time Use of the electromyogram filter should depend on signal conditions When there is a lot of noise in the output it is recommended that the LPF filter be set to 40Hz QRS Sound Setup...

Page 65: ...rom the patient s pacemaker is registered by Cardio M Plus users can set whether to print location of the pacemaker device in rhythm and diagnostic printing modes or not In P M menu within the Device Setup users can set it to either ON or OFF ON for placing the pacemaker location on the printout and OFF for disabling the pacemaker location print Note We recommend that P M be set to OFF in normal c...

Page 66: ...the menu with ON turning on the message display and OFF switching off the message display The following message for Lead Fault will be shown with affected channels Note If the RL lead becomes faulty the device will display a message saying all leads are faulty Lead Fault All Leads If a Lead Fault occurs during monitoring or recording a buzzer alarm will sound and the message box will pop up No pac...

Page 67: ...ts Thin indicates thin lines Normal is regular thickness and Thick is bold lines Demo Mode Setting This menu puts Cardio M Plus in demo mode When it is set to ON 60bpm Sinus Normal Rhythm signal will be displayed on the device and DEMO sign will appear on the upper part of the LCD screen Users can test rhythm diagnostic copy printing functions as well as PC communication and other miscellaneous fu...

Page 68: ...68 83 Version 1 0 Cardio Cardio Cardio Cardio M Plus M Plus M Plus M Plus Operation Manual Operation Manual Operation Manual Operation Manual ...

Page 69: ...printing time by utilizing 5 second buffering Note When the ON setting is used you should operate by pressing the Diagnostic Printing button after the signal is stabilized The device will utilize data from 5 seconds prior to button pressing for the function which means diagnostic information and output printout will be negatively affected if the signal is unstable QikPrint Quick Printing setting o...

Page 70: ...o M Plus The IP Address field sets a unique IP address which is assigned to the device by the hospital s network provider In the DB server field users can enter the IP address of the PC Workstation where EKG data from the device will be transmitted to i e IP address of the PC where BMS Plus software that can receive EKG information from device and store has been installed For subnet mask and gatew...

Page 71: ...dio Cardio M Plus M Plus M Plus M Plus Operation Manual Operation Manual Operation Manual Operation Manual Hospital Setup Enter names of the hospital clinic and attending doctor which will be recorded and printed along with the EKG data ...

Page 72: ...EKG Main screen Pressing the NET ESC key will trigger the device to transmit data to the PC workstation when the Data Management screen appears For guides in managing acquired data you should refer to the 10 Data Management Screen section Data transmission to a PC can only occur while the server program on the PC is running Therefore please check to make sure the server program on the PC is in ope...

Page 73: ...on Currently Selected Data Previous Page Next Page Current Cursor Position Number Of All Data Previous Menu Shown As ECG or SPR Show Menu Prior To Entering Data Management Screen Display Information Of Highlighted Patient Delete Highlighted Data From The Device Print Highlighted Data When you move to the right or left with the rotary key at the first or last data data will be shown per page ...

Page 74: ...Data With the NET ESC key users can transmit saved data to a PC workstation Attempting connection to PC workstation SYSTEM MESSAGE DB Connecting Transmitting data to PC workstation SYSTEM MESSAGE Sending Saved Data Indicating that the data transmission is finished SYSTEM MESSAGE Transmission Complete System Message if a connection to the PC has failed SYSTEM MESSAGE Connection Error System Message...

Page 75: ... flash drive Copying data to a USB flash drive SYSTEM MESSAGE File Copying Indicating that the data copy is finished SYSTEM MESSAGE File Copy Success System message if a USB flash drive is not connected SYSTEM MESSAGE Please checking the device System messages the reasons when the data copy failed SYSTEM MESSAGE File Copy Failed File Busy SYSTEM MESSAGE File Copy Failed Memory Full SYSTEM MESSAGE ...

Page 76: ...des are attached to the patient to prevent them from contacting conductive parts or contact with ground WARNING Use the provided ECG patient cable for measuring ECG in the operating room The patient cable cannot be used for measuring respiration WARNING Do not use the Cardio M Plus in combination with any Electro Surgical ES equipment WARNING Users must use the same type of electrode or any other ...

Page 77: ...0 Cardio Cardio Cardio Cardio M Plus M Plus M Plus M Plus Operation Manual Operation Manual Operation Manual Operation Manual Chapter 4 System Management 1 Maintenance And Cleaning 2 Regular Check Up 3 Trouble Shooting ...

Page 78: ...Do not soak the device and its cables into any liquid or detergent Keep the machine and its cables away from any liquid Disposal Of Your Old Appliance 1 All electrical and electronic products should be disposed separately from the municipal waste stream via designated collection facilities appointed by the government or the local authorities 2 Correct disposal of your old appliance will help preve...

Page 79: ...d to the signal Check whether the AC power supply filter is set up properly If it is still noisy even though the AC power supply filter is ON connect the ground wire to the GND electrode of the system Do not use the GND with that of AC power supply Instead connect GND to the patient s bed or any other metal around the device that is connected to the building s wall If you continue to have problems...

Page 80: ... screen alphanumeric keyboard and keypad push button and rotary switch Patient data ID name age sex height weight race smoke Basic measurement Heart rate PR QRS QT QTc P R T axis Recorder Thermal print head roll paper Report paper width A4 210mm or 8 5 length A4 300mm or 11 Resolution vertical 8dot mm Horizontal 16dot mm Electrical Resolution 500 sample sec Internal noise 20 µV p p max Input imped...

Page 81: ...normal use approx 100 ECG printouts Communication PC connection with RS232 interface and LAN Safety conformity ClassⅠ Type BF Environment Operating humidity 30 85 Operating temperature 10 40 50 to 104 Atmospheric pressure 70 106KPa Standard accessory Power cord 1 ea patient cable 1 ea limb 1 set 4 ea chest ball 1 set 6 ea chart paper 1 ea operation manual 1 ea Options Rechargeable battery cart han...

Page 82: ...me Electrocardiograph Model Name Cardio M Plus License Number License Date Serial Number Warranty period 2 years from the date of purchase Date of purchase Customer Hospital Name Address Contact No Seller s Name Manufacturer s Name Thanks for purchasing Cardio M Plus This product is a medical machine This product meets the strict quality requirements thoroughly ...

Page 83: ...o M Plus M Plus M Plus M Plus Operation Manual Operation Manual Operation Manual Operation Manual International Sales Service Medical Econet GmbH Mainstr 6 c d 45768 Marl Germany www medical econet com Medical Econet GmbH Model Name Cardio M Plus ...