MMF/SMF:
Selects the type of connector/fiber to be analyzed and loads the correct Pass/Fail mask. Please note that the label in the
button DOES NOT indicate the current mask type. It indicates that the mask would be changed to if activated.
Shake ON/OFF:
Turning the Shake ON may help in situations when the image is not very stable, such as inspecting a female
connector or bulkhead.
7.4.1.8 Captured Files Tab
This section of the user interface lists all the connectors’ end face images that have been captured with the fiber inspection
microscope. Users can and “Before” and “After” identifiers or remainders when comparing images before or after cleaning or polishing.
Clear All:
Deletes all the image files in the Captured Files list
Delete:
Deletes any selected files from the Captured Files list. Use the check-boxes to select
Compare:
Compares two images, for example, ones taken before and after cleaning. Use the check-boxes to select any two image
files and then tap on the Compare soft button. Use the Change Mode button on the comparison window to enable and disable the
analysis overlay.
MTTplus e_Manual D07-00-096P RevA00
Page 50 of 66
Содержание MTTplus
Страница 1: ...MTTplus e_Manual D07 00 096P RevA00 Page 1 of 66...
Страница 7: ...3 1 MTTplus Overview Side View Go back to TOC MTTplus e_Manual D07 00 096P RevA00 Page 7 of 66...
Страница 21: ...Web Browser Password Prompt MTTplus e_Manual D07 00 096P RevA00 Page 21 of 66...
Страница 40: ...Go back to TOC MTTplus e_Manual D07 00 096P RevA00 Page 40 of 66...
Страница 46: ...Go back to TOC MTTplus e_Manual D07 00 096P RevA00 Page 46 of 66...