Vectra Genisys® Therapy System
141
TrOuBLEShOOTiNg
ErrOr COdES (CONTiNuEd)
code
number
Type
Message
Probable cause
Possible remedies
108
Warning
Attempting to save additional User Protocols or Sequences after
system memory has reached the maximum allowed (200).
Delete some User Protocols or Sequences. Refer to appropriate section of the Therapy System User
Manual for instructions.
109
110
111
Warning
Warning
Warning
Attempting to access protocols or sequences and none are found
in the system.
a.
User Protocols- No protocols have been saved in the system. Refer to Therapy System User
Manual to save User Protocols.
b.
Sequences- No User Sequences have been saved in the system. Refer to Therapy System User
Manual to save Sequences.
112
Warning
Ultrasound Applicator disconnected from system during treatment
session.
a.
Connect Ultrasound Applicator to system.
b.
If Ultrasound Applicator is connected, reset system by turning power switch Off and On.
c.
If problem persists, connect a known good Ultrasound Applicator. If problem continues,
contact dealer or DJO, LLC for service.
113
Warning
Attempting to perform Ultrasound treatment with no Applicator
connected to the system.
a.
Connect the desired Ultrasound Applicator to the system.
b.
If Ultrasound Applicator is connected, reset system by turning power switch Off and On.
c.
If problem persists, connect a known good Ultrasound Applicator. If problem continues,
contact dealer or DJO, LLC for service.
114
Warning
Ultrasound Applicator is not calibrated.
Attempt to use a known good Applicator. If problem continues, contact dealer or DJO, LLC for
service.
115
Warning
Ultrasound Applicator is too hot.
Allow Ultrasound Applicator Sound Head to cool to ambient temperature.
116
117
Warning
Warning
1.
No Patient Data Card is inserted into the system.
2.
Attempted to use an Invalid Patient Data Card.
a.
Properly insert the Patient Data Card into the system port. Refer to Therapy System User
Manual for new and existing Patient Data Card instructions.
b.
Attempt to use a known good Patient Data Card.
c.
Make certain a Patient Data Card and not an sEMG Data Card is being used.
D.
If problem continues, contact dealer or DJO, LLC for service.