Vectra Genisys® Therapy System
75
OPErATiON
Refer to page 45
for Stim Editing
procedures .
After editing is complete, press the Back
button .
NOTE:
Stim treatment time is fixed and
cannot be changed .
s
EMg + STiM ThErAPY SET uP (CONTiNuEd)
TarGET
buTTOn
caPTurE TarGET
buTTOn
bEGIn caPTurE
buTTOn
Edit Stim
Press the Edit Stim button .
EDIT STIM buTTOn
back
buTTOn
Three options are available for setting the
sEMG modality target:
Max-
Maximum of muscle contractions in
10 seconds .
Avg-
Average of maximums achieved in
15 seconds of muscle contractions .
Manual-
Set Target manually .
Press the Target button until the prescribed
target option is displayed in the Target icon .
Select Target Option
Setting Max Target
Make certain Target Max is displayed in the
Target icon .
Press the Capture Target button . Have the
patient begin contracting the muscle and
press the Begin Capture button . Patient
should contract the muscle as hard as
possible during the preset 10 second
contraction period .
NOTE:
The capture may be stopped by
pressing the End Capture button .
The System will then select the
maximum contraction level
achieved during the preset
10 second contraction period .
Содержание Genisys 2761
Страница 158: ...ISO 13485 Certified 27190_H DJO LLC 2009 DJO LLC...