Vectra Genisys® Therapy System
129
iNSTALLATiON/rEMOvAL
The sEMG Plug Kit, part number 28027,
is designed to protect the exposed PC
Board contacts on the bottom of the
Therapy System and the Channel 3/4
Electrotherapy Module when the sEMG
Module has been removed .
The sEMG Plug Kit must be installed when
a sEMG Module has been removed for
service related problems and the Therapy
System will be put back into service
without the sEMG module .
The sEMG Plug Kit is installed and
removed in the same fashion as a sEMG
Module . Refer to
page 127
for installation
instructions .
Prepare the Therapy System for removal of
the sEMG Module . Refer to
Page 118
.
rEMOviNg
s
EMg MOduLE
Prepare System
Disconnect the system from the power
source (outlet or remove battery module if
installed) before attempting any maintenance,
installation, removal, or replacement
procedures to prevent electrical shock and
possible damage to system .
Remove sEMG Module
Insert a Flat Blade Screwdriver into the
Release Slot and pry against the Locking
Tab of the sEMG Module until the sEMG
Module is free from the Therapy System .
lOckInG Tab
rElEaSE SlOT
sEMG Plug Kit
The sEMG Module should only be removed for service related problems .