Vectra Genisys® Therapy System
138
OPTiON OPErATiON
When charging Battery Module, make
certain the System is electrically grounded
by connecting only to a grounded
electrical service receptacle conforming to
the applicable national and local electrical
codes . Using any other power source
will cause extensive damage to internal
components and render the System
unsafe for patient therapy .
ChArgiNg BATTErY MOduLE
Use the following instructions to properly
charge the Genisys Battery Module for use .
With Battery Module installed on the Vectra
Genisys Therapy System, plug the System
into an approved, grounded power outlet .
Turn the System on . The Lightning Bolt
icon will display indicating the battery
is charging . As the battery charges, the
Battery indicator icon will progressively fill
until it is solid black indicating full charge .
baTTEry
InDIcaTOr
IcOn
Charge the battery according to the
•
instructions found in this manual .
Never attempt to charge the battery
on any other charging mechanism .
Use the Battery Module only with the
•
Vectra Genisys Therapy System .
Do not reverse the polarity of the
•
Battery Module . Doing so can increase
the individual cell temperature and
cause cell rupture or leakage .
lIGhTnInG
bOlT IcOn
Charging efficiency is optimum within a
•
temperature range of 50 °F to 86 °F (10
°C to 30 °C) . Charge the Battery Module
within this temperature range .
At temperatures below 32 °F (0 °C) the
•
gas absorption reaction is not sufficient
and causes an increase of the gas
pressure inside the battery . This condition
can activate the safety vent and lead
to alkaline gas leaking and battery
performance deterioration .
Charging efficiency of the Battery Module
•
drops at temperatures above 104 °F
(40 °C) and can disrupt full charging,
lead to deterioration in performance and
battery cell leakage .
Charging of the Battery Module is required
when the Battery Indicator begins flashing .
When to Recharge
NOTE:
The Lightening Bolt icon will
continue to display as long as
the System is connected to
a power source .
Charging Temperature