13. Electromagnetic Compatibility (EMC) Information
54
EzRay Premium User Manual
Guidance and manufacturer’s declaration - electromagnetic immunity
The EzRay Premium is intended for use in the electromagnetic environment
specified below. The customer or the user of the EzRay Premium should assure that
it is used in such an electromagnetic environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Conducted
RF IEC
61000-4-6
Radiated RF
IEC 61000-4-
3
3 Vrms 150
kHz to
80MHz 3 V /
m 80 MHz to
2.5GHz
3 Vrms 150
kHz to 80MHz
3 V / m 80
MHz to
2.5GHz
The EzRay Premium must be
used only in a shielded location
with a minimum RF shielding
effectiveness and, for each cable
that exits the shielded location, a
minimum RF filter attenuation of
20 dB from 30 MHz to 230 MHz,
20 dB from 230 MHz to 1 GHz.
Field strengths outside the
shielded location from fixed RF
transmitters, as determined by an
electromagnetic site survey,
should be less than 3 V / m.a
Interference may occur near
equipment marked with the
following symbol:
NOTE 1) These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects, and
people.
NOTE 2) It is essential that the actual shielding effectiveness and filter attenuation
of the shielded location be verified to assure that they meet the minimum
specification.
Field strength from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength
outside the shielded location in which the EzRay Premium is used exceeds 3V/m,
the EzRay Premium should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as
relocating the EzRay Premium or using a shielded location with a higher RF
shielding effectiveness and filter attenuation.
Содержание EzRay Premium VEX-S100W
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...Notice vi EzRay Premium User Manual Left blank intentionally...
Страница 14: ...1 General and Regulatory Information 6 EzRay Premium User Manual Left blank intentionally...
Страница 24: ...3 The System Overview 16 EzRay Premium User Manual Left blank intentionally...
Страница 36: ...5 Using the System 28 EzRay Premium User Manual Left blank intentionally...
Страница 40: ...6 Emergency Measures 32 EzRay Premium User Manual Left blank intentionally...
Страница 46: ...9 Service 38 EzRay Premium User Manual Left blank intentionally...
Страница 48: ...10 Disposing of the Unit 40 EzRay Premium User Manual Left blank intentionally...
Страница 52: ...11 Technical Specifications 44 EzRay Premium User Manual...
Страница 58: ...12 X Ray Dose Data 50 EzRay Premium User Manual Measurement Values Longitudinal...
Страница 64: ......