12. X-Ray Dose Data
EzRay Premium User Manual
47
EN
G
L
ISH
12. X-Ray Dose Data
The X-ray dose data is extracted from the X-Ray Dose Test Report for the EzRay
Premium. The X-ray doses of EzRay Premium in the test report were measured in
accordance with the IEC collateral standards. EzRay Premium was designed in
accordance with Part 1. General Requirements for Safety, IEC 60601-1-3.
Test Condition
Brand Name (Model)
EzRay Premium (VEX-S100W)
Tube type
Toshiba D-041SB
Loading Factor
60 kV, 6.0 mA
Tube type / Inverter type
DG-10A05T3 / INV-05
12.1
X-Ray Dose Table
Test Equipment
Instrument
Manufacturer
Model
S/N
Dose Meter
Piranha
255
CB2-08050121
Dose Table
t (s)
Dose (mGy)
t (s)
Dose (mGy)
0.04
0.05
0.50
0.59
0.05
0.07
0.55
0.70
0.08
0.10
0.60
0.76
0.10
0.13
0.70
0.89
0.15
0.19
0.80
1.01
0.20
0.25
0.90
1.15
0.25
0.31
1.00
1.29
0.30
0.38
1.50
1.93
0.45
0.57
2.00
2.57
Содержание EzRay Premium VEX-S100W
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...Notice vi EzRay Premium User Manual Left blank intentionally...
Страница 14: ...1 General and Regulatory Information 6 EzRay Premium User Manual Left blank intentionally...
Страница 24: ...3 The System Overview 16 EzRay Premium User Manual Left blank intentionally...
Страница 36: ...5 Using the System 28 EzRay Premium User Manual Left blank intentionally...
Страница 40: ...6 Emergency Measures 32 EzRay Premium User Manual Left blank intentionally...
Страница 46: ...9 Service 38 EzRay Premium User Manual Left blank intentionally...
Страница 48: ...10 Disposing of the Unit 40 EzRay Premium User Manual Left blank intentionally...
Страница 52: ...11 Technical Specifications 44 EzRay Premium User Manual...
Страница 58: ...12 X Ray Dose Data 50 EzRay Premium User Manual Measurement Values Longitudinal...
Страница 64: ......