5. Using the System
24
EzRay Premium User Manual
Bitewing
For a bitewing exposure, the patient closes their teeth during exposure on the
film’s tab or on the film/sensor holder.
The x-ray beam is directed downward at 5° ~ 8°.
Teeth
Angle of inclination
Bitewing exposure
+5
˚
~ +8
˚
Positioning the Imaging Receptor
Using the EzRay Premium Intra-Oral X-ray system, you may create an X-ray image on
one of three different types of imaging receptors:
Conventional dental films
Digital sensors, such as VATECH’s EzSensor
Phosphor plate
Placing the receptor properly is important. Check your dental radiography text for
information about the proper placement of the imaging receptor. Improperly positioning
the film or sensor results in errors on the radiograph, such as distorted teeth and roots,
elongation, magnification, and/or overlapping contacts.
The paralleling technique reduces the risk of such errors. However, if you improperly
position the receptor, angulation errors can occur (angulation of the receptor to the
tooth itself).
If the exit pattern of the beam is not aligned with the imaging receptor, then part of the
radiograph will not be exposed to radiation and the final radiograph will have some
clear (unexposed) areas.
Содержание EzRay Premium VEX-S100W
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...Notice vi EzRay Premium User Manual Left blank intentionally...
Страница 14: ...1 General and Regulatory Information 6 EzRay Premium User Manual Left blank intentionally...
Страница 24: ...3 The System Overview 16 EzRay Premium User Manual Left blank intentionally...
Страница 36: ...5 Using the System 28 EzRay Premium User Manual Left blank intentionally...
Страница 40: ...6 Emergency Measures 32 EzRay Premium User Manual Left blank intentionally...
Страница 46: ...9 Service 38 EzRay Premium User Manual Left blank intentionally...
Страница 48: ...10 Disposing of the Unit 40 EzRay Premium User Manual Left blank intentionally...
Страница 52: ...11 Technical Specifications 44 EzRay Premium User Manual...
Страница 58: ...12 X Ray Dose Data 50 EzRay Premium User Manual Measurement Values Longitudinal...
Страница 64: ......