5. Using the System
26
EzRay Premium User Manual
5.4
Tables of Exposure Times (Default)
The following exposure timetables were established with a unit equipped with a cone
that corresponds to a focus-to-skin distance of 20 cm (8 inches) and 30 cm (12 inches)
respectively.
Receptor
Adult
Standard:
Cone 20 cm
(8 inches)
Option:
Cone 30 cm
(12 inches)
kV
mA
sec
kV
mA
sec
FILM
Incisor
60
7
0.08
70
6
0.13
Premolar
60
7
0.12
70
6
0.18
Molar
60
7
0.17
70
6
0.25
Bitewing
60
7
0.18
70
6
0.27
SENSOR
Incisor
60
7
0.07
70
6
0.11
Premolar
60
7
0.10
70
6
0.16
Molar
60
7
0.13
70
6
0.20
Bitewing
60
7
0.14
70
6
0.21
Receptor
Child
Standard:
Cone 20 cm
(8 inches)
Option:
Cone 30 cm
(12 inches)
kV
mA
sec
kV
mA
sec
FILM
Incisor
60
7
0.05
70
6
0.08
Premolar
60
7
0.07
70
6
0.11
Molar
60
7
0.11
70
6
0.14
Bitewing
60
7
0.11
70
6
0.14
SENSOR
Incisor
60
7
0.04
70
6
0.07
Premolar
60
7
0.06
70
6
0.10
Molar
60
7
0.08
70
6
0.13
Bitewing
60
7
0.09
70
6
0.14
Содержание EzRay Premium VEX-S100W
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...Notice vi EzRay Premium User Manual Left blank intentionally...
Страница 14: ...1 General and Regulatory Information 6 EzRay Premium User Manual Left blank intentionally...
Страница 24: ...3 The System Overview 16 EzRay Premium User Manual Left blank intentionally...
Страница 36: ...5 Using the System 28 EzRay Premium User Manual Left blank intentionally...
Страница 40: ...6 Emergency Measures 32 EzRay Premium User Manual Left blank intentionally...
Страница 46: ...9 Service 38 EzRay Premium User Manual Left blank intentionally...
Страница 48: ...10 Disposing of the Unit 40 EzRay Premium User Manual Left blank intentionally...
Страница 52: ...11 Technical Specifications 44 EzRay Premium User Manual...
Страница 58: ...12 X Ray Dose Data 50 EzRay Premium User Manual Measurement Values Longitudinal...
Страница 64: ......