5. Using the System
EzRay Premium User Manual
27
EN
G
L
ISH
5.5
X-Ray Generation Limit
Tube Voltage, Tube Current, and Exposure Time value cannot be set
to the maximum value simultaneously.
If the parameter you set exceeds the safety limit, the “Over Value”
message is displayed on the top left of the LCD. Unless this message
is cleared, you cannot perform irradiation.
Tube Voltage (kV)
Tube Current (mA)
Exposure Time (sec)
60 ~ 65
4 ~ 6
0.04 ~ 2.0
61 ~ 70
7
Not available
66 ~ 70
4 ~ 5
0.04 ~ 2.0
60
7
0.04 ~ 0.5
66
6
0.04 ~ 1.8
67
6
0.04 ~ 1.5
68
6
0.04 ~ 1.2
69
6
0.04 ~ 0.9
70
6
0.04 ~ 0.5
Adjustable step: 1 kV, 1 mA, 0.01 sec
Содержание EzRay Premium VEX-S100W
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...Notice vi EzRay Premium User Manual Left blank intentionally...
Страница 14: ...1 General and Regulatory Information 6 EzRay Premium User Manual Left blank intentionally...
Страница 24: ...3 The System Overview 16 EzRay Premium User Manual Left blank intentionally...
Страница 36: ...5 Using the System 28 EzRay Premium User Manual Left blank intentionally...
Страница 40: ...6 Emergency Measures 32 EzRay Premium User Manual Left blank intentionally...
Страница 46: ...9 Service 38 EzRay Premium User Manual Left blank intentionally...
Страница 48: ...10 Disposing of the Unit 40 EzRay Premium User Manual Left blank intentionally...
Страница 52: ...11 Technical Specifications 44 EzRay Premium User Manual...
Страница 58: ...12 X Ray Dose Data 50 EzRay Premium User Manual Measurement Values Longitudinal...
Страница 64: ......