Conventions Used in this Guide
viii
EzRay Air W (VEX-S300W) Installation Manual
Installation site
Proper shielding of the room is essential: Since these
requirements vary depending on the country, it is the
installer’s responsibilities to verify that all applicable radiation
safety requirements are met.
This equipment should not be installed in the immediate
vicinity of other devices.
Do not install the equipment in an area that is exposed to
strong electromagnetic fields.
Do not install this system in an area where there is the risk of
an explosion.
The electrical installation of this system shall comply with all
local code requirements for electro-medical systems:
IEC
60364-7-710:2002.
It is strongly recommended that a UPS be installed at the
same time as the equipment.
This equipment complies with the following standards:
IEC 60601-1
IEC
/
EN 60601-1-2
IEC 60601-1-3
IEC 60601-1-6
IEC 60601-2-65
ISO 13485
21 CFR 1020.30 & 1020.31
Содержание EzRay Air W
Страница 1: ......
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 10: ...Table of Contents x EzRay Air W VEX S300W Installation Manual This page intentionally left blank...
Страница 12: ...1 Introduction 2 EzRay Air W Model VEX S300W Installation Manual This page intentionally left blank...
Страница 54: ...4 Product Specifications 44 EzRay Air W Model VEX S300W Installation Manual...
Страница 60: ...4 Product Specifications 50 EzRay Air W Model VEX S300W Installation Manual This page intentionally left blank...
Страница 61: ...Appendix EzRay Air W Model VEX S300W Installation Manual 51 ENGLISH Appendix Drilling Template...
Страница 64: ...Appendix 54 EzRay Air W Model VEX S300W Installation Manual This page intentionally left blank...
Страница 66: ......
Страница 67: ......
Страница 68: ......