3.2.3 Leveling the Power Box Assembly: Front/Rear
1. Place the spirit level on the position as shown in the figure.
2. To adjust the horizontality (level), loosen one wrench bolt by using the wrench.
3. Adjust the headless wrench bolt until the bubble on the spirit level is placed in the
middle.
Allen wrench
5 mm
Since the Power Box Assembly is shipped with one bolt tightened, they
are required to be loosened prior to leveling it.