8. Acquiring CEPH Images (Optional)
vatech A9 (Model name: PHT-30CSS) User Manual
91
Englis
h
8.3.4 Waters’ view Mode
Patient Positioning
1.
Turn the Nasal Positioner to the
PA/Waters’ view/Carpus
mode Positioning
Marker, as shown below.
2.
Fold the Nasal Positioner up. The Nasal Positioner is not used in Waters’ view
mode.
3.
Guide the patient to the CEPH unit.
4.
Ask the patient to sit upright facing the sensor. Make sure that the patient’s
shoulders are level and that his/her neck is relaxed.
5.
Adjust the instrument to the patient's sitting height using the Column UP/DOWN
on the Handle frame or the optional UP/DOWN switch.
After adjusting the height of the column, align the Ear Rods
to the patient.
6.
During the operation, correctly align the Ear Rods to the patient's ears, so his/her
head does not move.
7.
Direct the patient to swallow first before closing his/her mouth and guide the
patient to bend the head backward 30˚ - 40˚. Direct the patient to remain in the
current position until image acquisition is completed.
Содержание A9
Страница 1: ......
Страница 2: ...Notice 2 vatech A9 Model name PHT 30CSS User Manual...
Страница 4: ...Notice 4 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Страница 12: ...1 Introduction 12 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Страница 18: ...2 General Information 18 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Страница 124: ...10 Troubleshooting 124 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Страница 140: ...12 Maintenance 140 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Страница 142: ...13 Disposing of the Equipment 142 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Страница 154: ...14 Technical Specifications 154 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Страница 170: ...15 Appendices 170 vatech A9 Model name PHT 30CSS User Manual...
Страница 194: ......
Страница 196: ......