12. Maintenance
vatech A9 (Model name: PHT-30CSS) User Manual
139
Englis
h
12.4.3.2 Analyzing S&C Check Phantom
1.
Remove Bite and Temple Supports from the Unit.
2.
Double-click the saved PHANTOM image in EzDent-i to run Ez3D-i.
3.
Go to the Axial view and double-click the pane for full-screen mode.
4.
Adjust the contrast by using WINDOWING Icon.
5.
Adjust the Brightness by using WINDOWING Icon.
6.
Make sure that you can see the minimum size of the Airhole and the PE
(Polyethylene) cylinder at a distance of 50 cm (20 inches) from the monitor.
Parameters
Material
Minimum visible size
High Contrast Resolution
Air Hole
1.0mm
Low Contrast Resolution PE cylinder
8.0mm
<High Contrast Resolution>
<Low Contrast Resolution>
Содержание A9
Страница 1: ......
Страница 2: ...Notice 2 vatech A9 Model name PHT 30CSS User Manual...
Страница 4: ...Notice 4 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Страница 12: ...1 Introduction 12 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Страница 18: ...2 General Information 18 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Страница 124: ...10 Troubleshooting 124 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Страница 140: ...12 Maintenance 140 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Страница 142: ...13 Disposing of the Equipment 142 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Страница 154: ...14 Technical Specifications 154 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Страница 170: ...15 Appendices 170 vatech A9 Model name PHT 30CSS User Manual...
Страница 194: ......
Страница 196: ......