8. Acquiring CEPH Images (Optional)
82
vatech A9 (Model name: PHT-30CSS) User Manual
5.
Select X-ray intensity.
Depending on the circumference of the patient’s head, X-ray
intensity may be classified as Hard, Normal, or Soft:
Soft ≤ Normal ≤ Hard
6.
The values of tube voltage and current are configured automatically according to the
patient's gender/age group and X-ray intensity. Click the
UP/DOWN
arrow to adjust
kVp and mA. The dose is adjustable by ±1 kVp and ±1 mA, respectively.
7.
Click the
CONFIRM
button when the exposure parameter setting is completed.
When you click
CONFIRM
button,
The Rotating Unit will move to its initial scanning
position.
The Vertical Beam will be activated to make patient
positioning easier.
The DAP (Dose Area Product), Scan Time, and
Exposure Time will be displayed below the Scan
Information window.
8.
Guide the patient to the equipment.
Содержание A9
Страница 1: ......
Страница 2: ...Notice 2 vatech A9 Model name PHT 30CSS User Manual...
Страница 4: ...Notice 4 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Страница 12: ...1 Introduction 12 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Страница 18: ...2 General Information 18 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Страница 124: ...10 Troubleshooting 124 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Страница 140: ...12 Maintenance 140 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Страница 142: ...13 Disposing of the Equipment 142 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Страница 154: ...14 Technical Specifications 154 vatech A9 Model name PHT 30CSS User Manual Left blank intentionally...
Страница 170: ...15 Appendices 170 vatech A9 Model name PHT 30CSS User Manual...
Страница 194: ......
Страница 196: ......