12. Maintenance
vatech A9 (Model name: PHT-30CSS) User Manual
133
Englis
h
12.4.1.2 Imaging CT Number Phantom
2. Choose the
CBCT
button in the Console Software.
3. Select an item below.
Item
Value
FOV (Diameter x Height)
8 x 8
Image Option
High Resolution
Voxel Size (0.20)
Standard
Gender/Age group
Man
X-ray Intensity
Hard
4. Click the
Confirm
button.
5. Click the
READY
button when enabled.
6. Capture the PHANTOM image according to Console Software instructions.
7. When image capturing is completed, save the image in EzDent-i.