ProStar 210/215/ PrepStar 218/SD-1 SDM
-23-
03-914947-20:5
In the
Acquisition parameters
tab, required fields like the
Run Name
, the
Run ID
and the
Acquisition length
must be
filled in before starting an analysis. If the system is equipped
with an autosampler controlled by the Galaxie
Chromatography Data System, the
Injection Volume
, the
Vial
position and the
Rack
number (if required) must be entered.
Other parameters for each acquisition channel can be
entered in the
Channel X
tab.
Click on the
Start
button to start the acquisition if the system
is a single-injector system or click on the
Next
button if the
system is a multiple injector system. In a multiple injector
system, the same parameters as the ones entered for the
first injector must be entered before starting the acquisition.
Sequence
1. To create a sequence, select the
FILE / NEW SEQUENCE
menu, complete the displayed creation wizard screen (refer
to Galaxie Chromatography Data System User’s Guide), by
entering the system name that will perform the analysis, the