Injection Control
This window becomes active only when a pump module is
configured as an injector in the system (for more details refer to
the
Configuration of the ProStar / PrepStar pump
chapter).
Thru Pump Injection Parameter Group:
Enter the time when the injection will start in the
Injection time
field.
Enter the flow rate of the injection in the
Injection Rate
field. The
end time of the injection is determined according to the flow rate
and the injection volume entered in the Acquisition method
section (or in the Quick Start window or in the Sequence).
NOTE:
The
Injection Volume
entered in the Acquisition part of the method must be
entered in mL only.