ProStar 210/215/ PrepStar 218/SD-1 SDM
-19-
03-914947-20:5
Pressure Control
Pressure Limits Parameter Group:
Enter the
minimum and maximum pressure limits
of the pump.
Outside these limits, the pump will stop and an alarm will sound.
NOTE:
- The maximum pressure limit of a ProStar / PrepStar module depends on the type of its
pump head. This limit is automatically adapted according to the connected pump head.
To know these limits, refer to the Pressure Limits chapter in the appendix of this manual.
- If a higher value of pressure is entered by error, a message appears indicating the
pressure limits for the connected pump head and the last entered valid value is
displayed.
- If several ProStar / PrepStar modules are connected (in case of binary, ternary or
quaternary pump), the maximum pressure limit is the lowest value among the different
pump ones.