Models 947, 948 and 960 Leak Detectors Operations and Service Manual
3-12
Table 3-1 lists the Not Ready condition codes and the action to take to correct the error
condition.
3.4.1 Calibration and Sensitivity Checks
To perform calibration and sensitivity checks:
1. Warm up the 947/948/960 for 30 minutes.
2. Place the test port plug in the test port opening
➅
. Refer to Figure 3-2 on page 3-9.
3. Set the Transfer switch to
AUTO
.
4. Set the Mode switch to
AUTO
. The VENT LED should illuminate.
Table 3-1
NOT READY Condition Codes
Code
Problem Solution
c0
Amplifier Offset is
out of range
Notify Varian Vacuum Technologies Customer Service. Spectrometer
Tube Amplifier needs service.
c2
Rough pump is off
Turn on the rough pump.
c4
No emission
Select the other ion source filament. Replace ion source if necessary.
Refer to Section 4, “Maintenance”.
c5
Auto calibrate function is
out of range
Notify Varian Vacuum Technologies Customer Service. Leak detector
requires maintenance.
c6
Auto tune function is
out of range
Tap the AUTO CAL button. (A brief tap may clear this code to allow
the leak detector to operate.) If this clears c6 then the leak detector is
operable. If c5 appears with c6 it is the priority condition code.
Press
VENT
to clear.
NOTE: A c6 condition code means that maintenance of the leak
detector may soon be necessary. Notify Varian Vacuum Technologies
Customer Service at 1-800-8VARIAN.
c7
Auto Zero function is
out of range
Notify Varian Vacuum Technologies Customer Service. Leak detector
requires maintenance.
c8
Power interruption
Perform a Calibration and Sensitivity Check. This condition code may
be accompanied by an intermittent c9 indication.
c9
System pressure is
too high
Check for vacuum leaks. Refer to Section 4, “Maintenance”.
Содержание 947
Страница 12: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 16: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 80: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 120: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 160: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual 4 40 ...
Страница 176: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 206: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 214: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 228: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 241: ......