Models 947, 948 and 960 Leak Detectors Operations and Service Manual
A-5
[SHIFT]
PRSRS
SHOW
VLV ST
Displays the Test Port and System Pressures.
SET
OFFSET
SHOW
Displays the value of the electronic offset in the Preamplifier.
[SHIFT]
SET
OFFSET
SHOW
Sets a temporary value of the electronic offset in the
Preamplifier. Enter XX (the desired value), followed by the SHIFT
key, then the function key.
INIT
RPLR
PUT
Sets a temporary new value for Repeller Voltage. Enter XXX
(the desired value), then the function key.
[SHIFT]
INIT
RPLR
PUT
Sets a permanent new value for Repeller Voltage. Enter XXX
(the desired value), followed by the SHIFT key, then the
function key
INIT
ION V
PUT
Sets a temporary new value for Ion Voltage. Enter XXX
(the desired value), then the function key.
[SHIFT]
INIT
ION V
PUT
Sets a permanent new value for Ion Voltage. Enter XXX
(the desired value), followed by the SHIFT key, then the function
key.
INIT
V FOC
PUT
Sets a temporary new value for Variable Focus Voltage. Enter XXX
(the desired value), then the function key.
[SHIFT]
INIT
V FOC
PUT
Sets a permanent new value for Variable Focus Voltage. Enter
XXX (the desired value), followed by the SHIFT key, then the
function key.
INIT
GAIN
PUT
Sets a temporary new value for System Gain. Enter X.X
(the desired value), then the function key.
[SHIFT]
INIT
GAIN
PUT
Sets a permanent new value for System Gain. Enter X.X
(the desired value), followed by the SHIFT key, then the
function key.
INIT
EMIS
PUT
Sets a temporary new value for Emission Current. Enter XXXX
(the desired value), then the function key.
Содержание 947
Страница 12: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 16: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 80: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 120: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 160: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual 4 40 ...
Страница 176: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 206: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 214: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 228: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 241: ......