Models 947, 948 and 960 Leak Detectors Operations and Service Manual
2-15
2.5 Operation using the Optional Rear Panel Interface
Once setup is complete and the 947/948/960 is warmed up, operation can be turned over
to the discrete I/O by asserting the PARALLEL ENABLE input (J120 pin 6). Operations can be
monitored using the front panel display, or via a PLC.
2.5.1 Parallel Enable
The PARALLEL ENABLE input turns control of the 947/948/960 over to the discrete I/O
for efficient operation. Once PARALLEL ENABLE is asserted, operations can be monitored
using the host serial port, but most parameter values cannot be changed. If a command is
issued that is not allowed once PARALLEL ENABLE is asserted, the system response is
cant
.
See Appendix C, “Communications Protocol”.
Operations can be monitored on the front panel display, but the parameter values cannot be
changed using the front panel. See Appendix D, “Inputs and Outputs”.
To remove control from the discrete I/O, you must de-assert PARALLEL ENABLE
(J120 pin 6).
NOTE
Control remains with the discrete I/O, even if the leak
detector is shut down while PARALLEL ENABLE is
asserted.
Содержание 947
Страница 12: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 16: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 80: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 120: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 160: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual 4 40 ...
Страница 176: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 206: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 214: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 228: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 241: ......