P3
– Faulty Combined PCB Air Switches
IMPORTANT
NOTE:
There are no serviceable parts on the Combined PCB. These air switches
are not serviceable and the whole board must be replaced if they fail.
As this is part of the
pneumatic system, it is important that all connections are airtight
.
Please note that all lifts will
come with new combined (Main + Aux.)PCB’s, C-450/C-625 Manual has only 2 air switches on
PCB, C-450/C-625 Power has 4 air switches on PCB and C-450/C-625 Power X-Y or Power
Turntable has 6 air switches on PCB.
1.
If there are no leaks in the hand control, airline tubing and grommets, there could be a leak
in air tubes to the pressure switch or failure of air switch. Check by performing the air leak
test with the hand control buttons.
2.
If there is an air leak,
tighten all air tubes and recheck.
Front side of new combined PCB (Main + Aux.) with air switches
Up
Down
Travers-
Gantry-
C-450/C-625 Technical Manual
031417
Page 15
Содержание C450
Страница 11: ...C 450 C 625 Technical Manual 031417 Page 11 ...
Страница 36: ...C 450 C 625 Technical Manual 031417 Page 36 ...
Страница 38: ...C 450 C 625 Technical Manual 031417 Page 38 ...
Страница 39: ...C 450 C 625 Technical Manual 031417 Page 39 ...
Страница 40: ...l C 450 C 62 Technical C 450 C 625 Technical Manual 031417 Page 40 ...
Страница 41: ...C 450 C 625 Technical Manual 031417 Page 41 ...
Страница 42: ...C 450 C 625 Technical Manual 031417 Page 42 ...
Страница 43: ...C 450 C 625 Technical Manual 031417 Page 43 ...
Страница 44: ...C 450 C 625 Technical Manual 031417 Page 44 ...
Страница 45: ...C 450 C 625 Technical Manual 031417 Page 45 ...
Страница 46: ...17 C 450 C 625 Technical Manual 031417 Page 46 ...
Страница 47: ...C 450 C 625 Technical Manual 031417 Page 47 ...