Electrical Systems
3.
If the batteries and circuit boards appear in order, no air leak is found and no fault is
found with the wiring
•
Proceed to check that the lift and traversing motors are functioning by removing the
motor and battery leads and running jumper cables from the motor leads to the battery
leads. (Reversing polarity causes the motors run in reverse)
4.
If the batteries and circuit boards appear in order, no air leak is found and no fault is
found with the wiring or with the motors
•
The most likely cause of failure is a blown fuse on the circuit board. To confirm, check
the fuses on the board and verify they are both operable. If a fuse is blown replace it
with another 15 amp fuse and test the lift.
Under no circumstances should a fuse with a
fuse rating other than 15 amps be used. Using such a fuse can result in damage to the
lift and /or personal injury.
If the problem persists after testing, replace the circuit board.
If the problem still persists contact customer service at 1-800-265-0677.
•
The other likely cause of failure is a circuit board failure that is not readily visible. To
confirm, connect the non-functioning lift to a board from service parts or from another
lift. (There is no need for a complete board installation: the replacement board used for
confirmation can be hooked up provisionally outside the lift.)
5.
If the on/off buttons (and/or emergency lowering buttons) on both the hand control
and the lift do not work
•
Check the Emergency Shut-Off with Pull Cord. If it has been activated during an
emergency, call customer service. If it has been activated accidentally push the button
towards the lift until you hear a click. This will allow you to operate the lift again. If the
lift still fails to operate examine the Emergency Shut-Off assembly to make sure it is not
obstructed and activates and deactivates the switch, repair as required. If the problem
persists, replace the Emergency Shut-Off switch.
C-450/C-625 Technical Manual
031417
Page 7
Содержание C450
Страница 11: ...C 450 C 625 Technical Manual 031417 Page 11 ...
Страница 36: ...C 450 C 625 Technical Manual 031417 Page 36 ...
Страница 38: ...C 450 C 625 Technical Manual 031417 Page 38 ...
Страница 39: ...C 450 C 625 Technical Manual 031417 Page 39 ...
Страница 40: ...l C 450 C 62 Technical C 450 C 625 Technical Manual 031417 Page 40 ...
Страница 41: ...C 450 C 625 Technical Manual 031417 Page 41 ...
Страница 42: ...C 450 C 625 Technical Manual 031417 Page 42 ...
Страница 43: ...C 450 C 625 Technical Manual 031417 Page 43 ...
Страница 44: ...C 450 C 625 Technical Manual 031417 Page 44 ...
Страница 45: ...C 450 C 625 Technical Manual 031417 Page 45 ...
Страница 46: ...17 C 450 C 625 Technical Manual 031417 Page 46 ...
Страница 47: ...C 450 C 625 Technical Manual 031417 Page 47 ...