3.
Select sample points (follow-up samples with the corresponding laboratory value)
to the graph to find the best correlation.
4.
Press
1st / 2nd degree calibration
.
5.
Press to accept the calibration.
13.4 Manual compensation
The sections above describe the compensation procedure using the graphic model from
Valmet BRIDGE. The necessary coefficients can also be entered manually for example
an Excel-based follow-up sample table.
Take follow-up samples to the table including attenuation, pressure and temperature. In
this example attenuation = Chemical compensation.
Table 3: Follow-up sample table
Measured Value [Cs%]
Lab Value [Cs%]
Difference [Cs%]
Attenuation [dB]
3
2.99
0.01
50
2.7
2.9
-0.2
55.7
2.5
2.84
-0.34
57,8
2.43
2.99
-0.56
60.3
2.2
3.1
-0.9
64.3
Advanced calibration | 56 (93)