Calibrating the LigaSure Generator
Setup, T
est
s, and Adjustment
s
LigaSure Vessel Sealing Generator Service Manual
6
-
19
Step 3 –
Adjust the Clock
A.
Refer to the following table for the commands to adjust the time.
B.
When the current time is correct, press Enter to proceed to the next step.
Step 4 –
Check and Adjust the Current Sense Gain
A.
Follow the instructions on the terminal.
B.
Connect a 10 ohm load to the bipolar output jacks, with one side passing
through a Pearson 411 current transformer connected to a Fluke 8920A RMS
voltmeter.
C.
Activate the generator in macrobipolar mode using the rear footswitch
receptacle.
D.
Referring to the following table for the correct commands, adjust the gain
until the meter reads 2.08 A.
Hour
Minute
Second
y: Hour + 1
u: 1
i: 1
Y: Hour + 10
U: 10
I: 10
b: Hour - 1
n: Minute –1
m: Second - 1
B: Hour - 10
N: Minute – 10
M Second - 10
I gain
i: gain + 1
I: gain + 10
m: gain - 1
M: gain - 10