Warnings, Cautions, and Notices
Service Personne
l Safety
LigaSure Vessel Sealing Generator Service Manual
1
-
5
Calibration
Cleaning
Caution
To avoid inadvertent coupling and/or shunting of RF currents around the resistor
elements, keep the resistors at least four inches (10.2 cm) away from any metal
surface including tabletops and other resistors. This is especially true if several
resistors are connected in series or parallel to obtain a specified value. Do not
allow the resistor bodies to touch each other.
Notice
After calibration, the generator will be ready to use only after you initiate the
internal self-test by turning the generator off, then on.
Calibrate the generator after you install a new battery. Calibration values are lost
when the battery is replaced.
Calibrate the generator after you install a new control board.
Calibrate the generator after you service, repair, or install new components, or
after you replace any generator board assembly.
Notice
Do not clean the generator with abrasive cleaning or disinfectant compounds,
solvents, or other materials that could scratch the panels or damage the
generator.