Performance Characteristics
Te
chnical S
pecifications
LigaSure Vessel Sealing Generator Service Manual
4
-
7
Input Power
Power Cord Specification
This unit was equipped from the factory with either a 110VAC hospital grade
NEMA 5-15 power cord or a 220VAC CEE7/7 power cord. Should the AC power
cord need to be replaced to match another plug configuration, the replacement
plug/cable/receptacle configuration must meet or exceed the following
specifications:
100-120 VAC
Cable - SJT16/3, IEC color code, maximum length 15 ft (5 m)
Plug - minimum 10 A - 125 VAC
Unit receptacle - IEC female, minimum 10 A - 125 VAC
120 Volt
240 Volt
Maximum power at nominal line
voltage
Idle: 35 VA
Bipolar: 360 VA
Seal: 480 VA
Maximum power at nominal line
voltage
Idle: 35 VA
Bipolar: 360 VA
Seal: 480 VA
Full regulation range
90 to 135 Vac
Full regulation range
186 to 264 Vac
Operating range
85 to 140 Vac
Operating range
170 to 264 Vac
Mains current maximum
Idle: 300 mA
rms
Bipolar: 3.0 A
rms
Seal: 4.0 A
rms
Mains current maximum
Idle: 300 mA
rms
Bipolar: 1.5 A
rms
Seal: 2.0 A
rms
Mains line frequency range (nominal)
50 to 60 Hz
Mains line frequency range (nominal)
50 to 60 Hz
Fuses (2) 4 A, 250 V, 3 AG, SLO-BLO
Fuses (2) 4 A, 250 V, 3 AG, SLO-BLO
1
1. The nominal operating power of the LigaSure generator is 480 VA for
a maximum operating current of 2.0 A. However, because the unit is
fused for 4.0 A, the worst-case power would be 880–960 VA (due to in-
put voltages of 220–240 Volts), as indicated by the labeling on the rear
panel of the generator.
Power plug
3-prong hospital grade connector
Power plug
3-prong locally approved connector