Responding to System Alarms
7-20
LigaSure Vessel Sealing Generator Service Manual
191
192
193
194
195
196
197
198
199
MBR_POWER_SETTINGS_CORRUPT
ED
manage_button_results has detected a
corruption of present power settings or
recall power settings that indicates
memory has been corrupted or MACRO/
BIPOLAR powers are greater than 95
watts.
MA_INVALID_AUDIO_STATE
manage_audio has detected an invalid
state value in the audio state variable.
MA_INVALID_MODE_VALUE
manage_audio has detected an invalid
mode value in the
curr_settings.curr_mode field.
SAT_INVALID_MODE_VALUE
select_audio_tone has detected an
invalid mode value in the
curr_settings.curr_mode field.
MA_AT_INVALID_ALERT_TYPE
manage_audio_alert_transition detected
an invalid value in the alert type variable
current alert.
CCD_CAL_DATA_CORRUPTED
check_cal_data detected an invalid CRC
in the dual port ram cal store; therefore,
dpram cal data has been corrupted.
MAIN_INVALID_STATE_VALUE
Main has detected an invalid state value
in the current_generator_state variable.
SNESR_INVALID_MESSAGE_TYPE
se_exec_normal_serial_result has
detected an invalid serial message type
stored in the pending_serial_cmd.type
field.
USL_INVALID_LED_MASK
update_smart_leds has detected an
invalid parameter (led_mask) value
passed to it.
Turn the power switch off (O) then on again (|).
If the error reappears, replace the control board.
Number
Description
Recommended Action