VALHALLA SCIENTIFIC, INC. 
2701C PRECISION VOLTAGE STANDARD 
 
CALIBRATION PROCEDURE

 

_____________________________________________________________ 
 
1. General

 

 
This section provides general maintenance information and a procedure for calibrating the 
2701C. The Model 2701C DC Voltage Calibrator should be calibrated on a routine basis 
(every 12 months is recommended) to ensure continued accuracy. 
 
For best results, the 2701C should be warmed-up with the covers in place at the calibration 
temperature for at least 2 hours before performing the calibration procedure that follows.   
 

2.  Required Test Equipment 

 
The following test equipment is required to calibrate the Model 2701C DC Voltage 
Calibrator: 
 
1)  A DC voltage measuring device capable of measuring the following ranges: 
 
  

volts 

 

±

0.5

µ

  

0.1 

volts 

±

0.5

µ

  

volt 

 

±

1

µ

  

10 

volts 

±

10

µ

  

100 

volts 

±

100

µ

  

1000 

volts 

±

1mV  

 

 
2)  A high quality low-thermal lead set must be used for maximum accuracy.  The 

recommended lead set is Valhalla Option SL-48 or its equivalent. 

 
3)  If the 2701C is equipped with the optional 120mA output, a 10

 

±

0.005% precision 

resistor is required.  Valhalla lead set “Option C” may be used to connect to the resistor. 

 

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Содержание 2701C

Страница 1: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Страница 2: ...rming the calibration procedure that follows 2 Required Test Equipment The following test equipment is required to calibrate the Model 2701C DC Voltage Calibrator 1 A DC voltage measuring device capab...

Страница 3: ...t of the display 3 200mV Range 000000C displays Adjustments to the output are made using the rightmost three knobs on the 2701C front panel The knob on the far right provides the least amount of chang...

Страница 4: ...be calibrated separately skip this step 120mA Mode 000 000C displays Connect the 10 resistor to the 2701C OUTPUT terminals Monitor the voltage drop across the resistor with the voltmeter Adjust the 27...

Страница 5: ...ssary to bring the voltmeter reading to 1 00000 volts 10 V Enter the adjustment The display should indicate End CAL Return the rear panel key switch to the OPERATE position This concludes the procedur...

Страница 6: ...to skip a step without changing the calibration data A sample command is shown below in HP Basic If you are not using HP Basic consult your operator s manual for the correct command syntax OUTPUT 715...

Страница 7: ...dirt or by other contaminants then it is recommended that the unit be completely overhauled at a Valhalla Scientific Service Center Valhalla Scientific Inc Service Center 8318 Miramar Mall San Diego...

Страница 8: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Отзывы: