4)

    2V Range, 0.00000C displays.

  Adjust the three knobs as required to bring the 

output to 0 

±

3

µ

V.  Enter the adjustment by pressing the STANDBY/OPERATE 

key. 

 
5)  

20V Range, 00.0000C displays.

  Make the required adjustments to bring the output 

to 0 

±

20

µ

V and enter the change. 

 
6)   

120V Range, 000.000C displays.

  Make the required adjustments to bring the 

output to 0 

±

150

µ

V and enter the change. 

 
7)  

1200V Range, 0000.00C displays.

  Make the required adjustment to bring the 

output to 0 

±

1.5mV and enter the change.  Note: This step is automatically skipped 

if the instrument is equipped with Option LNF. 

 
8) 

If Option IT-2 is installed, the 2701C proceeds to the 120mA Mode and the display 
indicates 000.000C.  If this option is not installed or if it will be calibrated 
separately, skip this step. 

 

 120mA 

Mode, 

000.000C displays.  

Connect the 10

 resistor to the 2701C 

OUTPUT terminals.  Monitor the voltage drop across the resistor with the voltmeter.  
Adjust the 2701C knobs as described above until the voltmeter reads 0 

5

µ

V.  Enter 

the adjustment and remove the connections. 

 
9) 

200mV Range, .100000C displays.  

Reconnect the voltmeter to the 2701C SENSE 

terminals, if required.  The knobs are weighted to affect the output as indicated 
below: 

 

 

1 0 0 0 0 0 C 

.

 0.25 ppm of range 

 

 

 

 

 

 

 

 

1 0 0 0 0

0  C 

.

 4 ppm of range 

 

 

 

 

 

 

 

 

1 0 0 0

0  0 C 

.

 64 ppm of range 

 

 

 

 

 

 

 

 

 
 

Make the required adjustments to bring the reading on the voltmeter to .1000000 
volts 

±

0.5

µ

V.  Enter the adjustment by pressing the STANDBY/OPERATE key. 

 
10) 

2V Range, 1.00000C displays.

  Make the required adjustments to bring the output 

to 1.000000 volts 

±

5

µ

V.  Enter the adjustment. 

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Содержание 2701C

Страница 1: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Страница 2: ...rming the calibration procedure that follows 2 Required Test Equipment The following test equipment is required to calibrate the Model 2701C DC Voltage Calibrator 1 A DC voltage measuring device capab...

Страница 3: ...t of the display 3 200mV Range 000000C displays Adjustments to the output are made using the rightmost three knobs on the 2701C front panel The knob on the far right provides the least amount of chang...

Страница 4: ...be calibrated separately skip this step 120mA Mode 000 000C displays Connect the 10 resistor to the 2701C OUTPUT terminals Monitor the voltage drop across the resistor with the voltmeter Adjust the 27...

Страница 5: ...ssary to bring the voltmeter reading to 1 00000 volts 10 V Enter the adjustment The display should indicate End CAL Return the rear panel key switch to the OPERATE position This concludes the procedur...

Страница 6: ...to skip a step without changing the calibration data A sample command is shown below in HP Basic If you are not using HP Basic consult your operator s manual for the correct command syntax OUTPUT 715...

Страница 7: ...dirt or by other contaminants then it is recommended that the unit be completely overhauled at a Valhalla Scientific Service Center Valhalla Scientific Inc Service Center 8318 Miramar Mall San Diego...

Страница 8: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Отзывы: