uvSVX and uvSVXe Operating Manual: MANUAL-857569191-11 | issue 1.7
Page | 33
© 2021
– Valeport Ltd
9 Care & Maintenance
The uvSVX is remarkably robust, being primarily constructed of titanium. The only maintenance
required, other than periodic recalibration as necessary or recommended at maximum of 2 years, is
to keep the instrument and sensors as clean as possible. The instrument should be rinsed with
freshwater when recovered and any debris or growth gently removed.
Do not attempt to remove the acetal Lock Screw Cap from the Pressure Module. If debris is caught
under the cover attempt to remove it under slowly flowing water,
use no tools
. If this fails contact
Valeport for further advice. If the sensor diaphragm is damaged in any way it will have to be
replaced.
The instrument should be stored in its transit case when not in use. It is not necessary to remove
the Interchangeable Pressure Sensor Module during storage.
Any damage to this diaphragm will render the Pressure Module warranty invalid
9.1 Calibration
The uvSVX may be returned to Valeport or one of its approved laboratories for recalibration:
Code
Description
#022
Sets the SV Calibration coefficients
this function is password protected
#023
Read the SV Calibration coefficients
#031
Sets the Calibration output mode for all readings
#031;0 = raw
#031;1 = calibrated
this function is password protected
#032
Read the Calibration mode
#085
Sets the User Pressure Calibration coefficients
this function is password protected
#086
Read the User Pressure Calibration coefficients
#087
Sets the User Temperature Calibration coefficients
this function is password protected
#088
Read the User Temperature Calibration coefficients
#138
Read the date for the last SV and Temperature Calibration
#203
Read the Pressure Sensor Gain Calibration
#204
Read the Pressure Sensor Offset Calibration
#223
Read the Pressure Sensor Calibration date
#501
Read the Calibration history for Temperature
#502
Reads the Calibration history for Sound Velocity
#510
Set the User Calibration activation status
#510;0 = User Calibration OFF
#510;1 = User Calibration ON (applied after conversion to feet or metres)
#511
Read User Calibration activation status
Sensors should be calibrated every 2 years