Section 6 | Data Output Formats
© 2021
– Valeport Ltd
Page | 24
6.4 SVX2 with Salinity
Available from Firmware version 0650741A7
Tab separated string. Terminated with ^t<CRLF>
Command: #082;9
Format:
cccc.ccc M/SEC dddd.ddd DBAR tttt.ttt C CCCC.CCC MS/CM SSSS.SSS PSU
Notes:
Field
Format
Sound Velocity
cccc.ccc
units
nnnn
Depth
dddd.ddd
units
n
Temperature (°C)
tttt.ttt
units
n
Conductivity
CCCC.CCC
units
MS/CM
Salinity
ssss.sss
units
PSU
Output String Examples
1483.576 M/SEC 0010.122 M 0021.291 C 0000.142 MS/CM 0000.124 PSU
1483.578 M/SEC 0010.122 M 0021.293 C 0000.140 MS/CM 0000.122 PSU
1483.578 M/SEC 0010.121 M 0021.284 C 0000.142 MS/CM 0000.124 PSU
6.5 SVX2 without Salinity
Available from Firmware version 0650741A7
Tab separated string. Terminated with ^t<CRLF>
Command:
#082;10
Format:
cccc.ccc M/SEC dddd.ddd DBAR tttt.ttt C CCCC.CCC MS/CM
Notes:
Field
Format
Sound Velocity
cccc.ccc
units
nnnn
Depth
dddd.ddd
units
n
Temperature (°C)
tttt.ttt
units
n
Conductivity
CCCC.CCC
units
MS/CM
Output String Examples
1483.576 M/SEC 0010.122 M 0021.291 C 0000.142 MS/CM
1483.578 M/SEC 0010.122 M 0021.293 C 0000.140 MS/CM
1483.578 M/SEC 0010.121 M 0021.284 C 0000.142 MS/CM