In push-button calibration, the probe outputs are by default relative humidity in
the range 0 … 100 %RH in CH1 and temperature in the range -40 … +60 °C (-40 …
+140 °F) in CH2.
6.2.1 Performing two-point humidity and temperature
adjustment
1. Carry out chemical purge, if available.
Chemical purge option (page 14)
2. Uncover the adjustment buttons on the probe by removing the protective cover and the
calibration seal.
3. Open the protective plug.
There are 3 buttons marked
-
,
□
, and
+
. There is also a 2-colored indicator LED. See
4. To change HMP155 to the adjustment mode, press the
ADJ
button and hold it down until
the green indicator LED lights up.
5. To perform the low humidity offset adjustment, remove the filter and insert the probe into
a measurement hole of the dry end reference chamber (for example, LiCl: 11 % RH).
Do not touch the adjustment buttons before the conditions have stabilized.
This takes approximately 30 minutes.
6. Using the
-
and
+
buttons, make sure the A
out
voltage is correct and press the
ADJ
button.
The green indicator LED turns off and back on.
7. Insert the probe into the high-end reference chamber (for example, NaCl: 75 % RH
chamber in the humidity calibrator HMK15) and perform the high humidity gain
adjustment by using the
-
and
+
buttons to make sure the A
out
voltage is correct. To finish
the relative humidity calibration and adjustment, press the
ADJ
button and the red
indicator LED lights up.
If you do not want to perform the temperature adjustment at this time, press
the
ADJ
button twice. The indicator LED turns off. After a reset, HMP155 exits
the adjustment mode. Otherwise, continue with the next step.
8. Insert the probe into a known reference temperature (if HMK15 humidity calibrator is not
used) and let the temperature reading stabilize.
Do not touch the adjustment buttons before the conditions have stabilized.
HMP155 User Guide
M210912EN-D
68
Содержание HUMICAP HMP155
Страница 1: ...M210912EN D User Guide Vaisala HUMICAP Humidity and Temperature Probe HMP155 ...
Страница 8: ...HMP155 User Guide M210912EN D 6 ...
Страница 90: ...7 7 3 Radiation shield dimensions Figure 17 DTR503 and DTR13 dimensions HMP155 User Guide M210912EN D 88 ...
Страница 94: ......
Страница 98: ...HMP155 User Guide M210912EN D 96 ...
Страница 100: ......
Страница 101: ......
Страница 102: ...www vaisala com ...