7. Technical data
7.1 Measuring specifications
Table 30 Humidity measurement performance
Property
Description/value
Sensor
HUMICAP
â
180R and 180R2 for typical
applications
HUMICAP
â
180C and 180R2C for applications
with chemical purge and/or warmed probe
Observation range
0 … 100 %RH
Response time at +20 °C (+68 °F) in still air with
sintered Teflon filter
63 %: 20 s
90 %: 60 s
Factory calibration uncertainty at +20 °C
(+68 °F)
±0.6 %RH (0 … 40 %RH)
±1.0 %RH (40 … 95 %RH)
Accuracy (including non-linearity, hysteresis, and repeatability)
At +15 … +25 °C (+59 … +77 °F)
±1 %RH (0 … 90 %RH)
±1.7 %RH (90 … 100 %RH)
At −20 … +40 °C (−4 … +104 °F)
±(1.0 + 0.008 × reading) %RH
At −40 … −20 °C (−40 … −4 °F)
±(1.2 + 0.012 × reading) %RH
At +40 … +60 °C (+104 … +140 °F)
±(1.2 + 0.012 × reading) %RH
At −60 … −40 °C (−76 … −40 °F)
±(1.4 + 0.032 × reading) %RH
1)
Defined as ±2 standard deviation limits. Small variations possible (see also the calibration
certificate).
Table 31 Temperature measurement performance
Description
Value
Sensor
Pt100 RTD element, Class F 0.1 IEC 60751
Observation range
−80 … +60 °C (−112 … +140 °F)
Response time for additional temperature probe
in 3 m/s (7 mph) air flow
63 %: < 20 s
90 %: < 35 s
Accuracy with voltage output
At −80 … +20 °C (−112 … +68 °F)
±(0.226 - 0.0028 × |temperature|) °C
HMP155 User Guide
M210912EN-D
80
Содержание HUMICAP HMP155
Страница 1: ...M210912EN D User Guide Vaisala HUMICAP Humidity and Temperature Probe HMP155 ...
Страница 8: ...HMP155 User Guide M210912EN D 6 ...
Страница 90: ...7 7 3 Radiation shield dimensions Figure 17 DTR503 and DTR13 dimensions HMP155 User Guide M210912EN D 88 ...
Страница 94: ......
Страница 98: ...HMP155 User Guide M210912EN D 96 ...
Страница 100: ......
Страница 101: ......
Страница 102: ...www vaisala com ...