In HUMICAP180C/180RC/R2C models, the temperature sensor is
integrated with the relative humidity sensor and Vaisala recommends sending
HUMICAP180C/180RC/R2C sensors to Vaisala Service.
If you change the sensor yourself, these instructions apply, except that before
removing the damaged sensor, you must unsolder the connections from the
temperature sensor pins. You must also solder the new temperature sensor
connections to the temperature sensor pins. Be careful when soldering the
sensor pins.
CAUTION!
1. Remove the filter from the probe.
Changing HMP155 probe filter (page 62)
2. After removing the filter, check the O-ring and change it, if necessary.
Figure 11 Changing INTERCAP/
HUMICAP180R/HUMICAPR2 sensors
1
Filter
2
O-ring
3
HUMICAP sensor
4
Pt100 temperature sensor
3. Remove the damaged sensor and insert a new one.
Handle the new sensor by the plastic socket. Do not touch the
sensor plate.
CAUTION!
4. After the sensor change, perform humidity calibration and adjustment as described in
Adjusting and calibrating relative humidity after sensor change (page 72)
5. Attach a new filter on the probe.