4. Press
ENTER
a few times to check if the reading is stabilized. Let the reading stabilize,
type the reference temperature after the question mark and press
ENTER
3 times.
Example of 2-point adjustment:
>ct
T : 18.6038 1. ref ?
T : 18.6068 1. ref ?
T : 18.6098 1. ref ? 19.0
Press any key when ready ...
T : 49.5176 2. ref ? 50.0
OK
>
5. Move the probe into another reference temperature and let the reading stabilize. Type the
reference temperature after the question mark and press
ENTER
.
OK
indicates that the adjustment has succeeded.
6. Type the calibration information (date and text) to the probe memory. See commands
CTEXT
and
CDATE
.
7. Reset the probe with the
RESET
command.
More information
‣
‣
6.3.3 One-point humidity and temperature adjustment
You can do one-point humidity and temperature adjustment with the following serial line
commands:
• For humidity adjustment:
CRH
and
FCRH
• For temperature adjustment:
CT
and
CTA
To make an adjustment, type the command, press SPACE, and type the reference value.
Example
Crh 20.0
OK
6.3.4 Adjusting and calibrating relative humidity after sensor
change
After sensor change, you must calibrate and adjust the relative humidity.
1. Connect HMP155 to a computer.
Serial line communication (page 36)
HMP155 User Guide
M210912EN-D
72
Содержание HUMICAP HMP155
Страница 1: ...M210912EN D User Guide Vaisala HUMICAP Humidity and Temperature Probe HMP155 ...
Страница 8: ...HMP155 User Guide M210912EN D 6 ...
Страница 90: ...7 7 3 Radiation shield dimensions Figure 17 DTR503 and DTR13 dimensions HMP155 User Guide M210912EN D 88 ...
Страница 94: ......
Страница 98: ...HMP155 User Guide M210912EN D 96 ...
Страница 100: ......
Страница 101: ......
Страница 102: ...www vaisala com ...